FDA Adverse Event Malfunction Summary report: N

BLOOD GLUCOSE SYSTEM

MDR report key: 5279146 · Received December 9, 2015

Report

Report Number
1052693-2015-02406
Event Type
Malfunction
Date Received
December 9, 2015
Date of Event
November 11, 2015
Report Date
December 9, 2015
Manufacturer
NIPRO DIAGNOSTICS, INC.
Product Code
NBW
PMA / PMN Number
K080641
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4) PRODUCT RETURNED AND EVALUATED WITH NO DEFECT FOUND. MOST LIKELY UNDERLYING ROOT CAUSE: STRIP ISSUE.

Description of Event or Problem · 1

CUSTOMER COMPLAINS OF HI RESULTS. CUSTOMER STATES THAT SHE FEELS WELL AND REQUIRES NO MEDICAL ATTENTION. THE CUSTOMER'S EXPECTED BLOOD RESULTS ARE 179-189MG/DL FASTING. VERIFIED THE STRIPS EXPIRED 5/31/2018. CUSTOMER CONFIRMS THE STRIPS HAVE BEEN PROPERLY STORED AND WERE FIRST OPENED (B)(6) 2015. CUSTOMER PERFORMED A BACK TO BACK BLOOD TEST AND OBTAINED 586MG/DL AND HI FASTING. REVIEWED METER MEMORY: 1:HI (B)(6) 2015 01:42:00 PM FASTING:YES. 2:HI (B)(6) 2015 01:39:00 PM FASTING:YES. 3:HI L (B)(6) 2015 12:14:00 PM FASTING:YES. 4:HI (B)(6) 2015 12:13:00 PM FASTING:YES. 5:HI (B)(6) 2015 12:12:00 PM FASTING:YES. CUSTOMER IS CONCERN WITH ALL THE HI BLOOD RESULTS SHE IS GETTING. CUSTOMER STATES THAT SHE IS FEELING FINE SHE FEELS NORMAL. CUSTOMER STATES THAT SHE WAS AT HER DR. OFFICE YESTERDAY AND EVERYTHING WAS FINE BUT THEY CHANGE HER MEDICINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
808748 BLOOD GLUCOSE SYSTEM NBW NIPRO DIAGNOSTICS, INC. TRUERESULT PS2356

Patients

Seq Age Sex Outcome Treatment
1 0 YR