FDA Adverse Event Malfunction Summary report: N

LEICA BIOSYSTEMS

MDR report key: 5273523 · Received December 4, 2015

Report

Report Number
3004859032-2015-00004
Event Type
Malfunction
Date Received
December 4, 2015
Date of Event
October 9, 2015
Report Date
November 9, 2015
Manufacturer
LEICA BIOSYSTEMS NEWCASTLE LTD
Product Code
NJT
Removal / Correction Number
1423337-16-11-2015-005-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINT WAS INITIALLY REVIEWED AND DETERMINED TO BE NON REPORTABLE AS THERE WAS NO OBVIOUS MALFUNCTION. IT WAS BELIEVED IT DID NOT WORK CORRECTLY DUE TO THE CUSTOMER USING A DIFFERENT PROTOCOL THAN WHAT IS STATED IN THE INFO FOR USE BOOKLET. THE RETAINED UNIT WAS ALSO TESTED AND FOUND TO WORK OPTIMALLY. THE CUSTOMER KIT WAS RETURNED AND TESTED, WHICH THEN LED TO THE UNDERSTANDING THE PRODUCT DID MALFUNCTION. AN INVESTIGATION WAS PERFORMED AND A RECALL OF THE AFFECTED PRODUCTS IS NOW UNDERWAY. THE ROOT CAUSE OF THE MALFUNCTION IS A MANUAL FILL OF THE PRODUCT HAS BEEN CONDUCTED WITH THE INCORRECT COMPONENT.

Additional Manufacturer Narrative · 1

A PRODUCT RECALL WAS INITIATED ON 11/16/2015, OF (B)(4) BATCHES OF (B)(4) BOND POLYMER REFINE RED DETECTION WHICH HAVE THE POTENTIAL TO BE AFFECTED BY THE SAME FAILURE MODE AS THIS REPORT.

Description of Event or Problem · 1

ON (B)(6) 2015, LEICA (B)(4) RECEIVED A COMPLAINT REGARDING DS9390 BOND POLYMER REFINE RED DETECTION BATCH 41919 TO STATE THERE WAS NO STAINING OBSERVED WHEN USING THIS PRODUCT. THE CUSTOMER WAS NOTED TO USE THE PRODUCT FOLLOWING A DIFFERENT PROTOCOL TO WHAT IS SPECIFIED IN THE INFO FOR USE BOOKLET. THE PRODUCT WAS RETURNED FOR ANALYSIS AND FOUND TO NOT PERFORM CORRECTLY WHEN TESTED USING THE CORRECT PROTOCOL. THE TESTING WAS CONCLUDED ON (B)(6) 2015. THIS IS THE DATE THE ROOT CAUSE WAS IDENTIFIED AS A MALFUNCTION WITH THE PRODUCT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
797020 LEICA BIOSYSTEMS BOND POLYMER REFINE RED DETECTION NJT LEICA BIOSYSTEMS NEWCASTLE LTD 41619

Patients

Seq Age Sex Outcome Treatment
1 UNK