FDA Adverse Event Malfunction Summary report: N

BLOOD GLUCOSE SYSTEM

MDR report key: 5268082 · Received December 4, 2015

Report

Report Number
1052693-2015-02369
Event Type
Malfunction
Date Received
December 4, 2015
Date of Event
November 8, 2015
Report Date
December 7, 2015
Manufacturer
NIPRO DIAGNOSTICS, INC.
Product Code
NBW
PMA / PMN Number
K080641
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PRODUCT RETURNED AND EVALUATED WITH NO DEFECT FOUND. MOST LIKELY UNDERLYING ROOT CAUSE: USER'S TEST STRIP HAD POOR FILL.

Additional Manufacturer Narrative · 1

(B)(4). PRODUCT RETURNED AND EVALUATED WITH NO DEFECT FOUND. MOST LIKELY UNDERLYING ROOT CAUSE: USER'S TEST STRIP HAD POOR FILL. TYPE OF REPORTABLE EVENT, INADVERTENTLY "DEATH" WAS SELECTED, THE CORRECT OPTION FOR THIS COMPLAINT IS "MALFUNCTION".

Description of Event or Problem · 1

CUSTOMER COMPLAINS OF LOW RESULTS. CUSTOMER STATES THAT HE FEELS WELL AND REQUIRES NO MEDICAL ATTENTION. THE CUSTOMER'S EXPECTED BLOOD RESULTS ARE 130-160MG/DL FASTING. VERIFIED THE STRIPS EXPIRED 3/17/2018. CUSTOMER CONFIRMS THE STRIPS HAVE BEEN PROPERLY STORED AND WERE FIRST OPENED (B)(6) 2015. CUSTOMER PERFORMED A BLOOD TEST AND OBTAINED 64MG/DL. REVIEWED METER MEMORY. 1:65MG/DL (B)(6) 2015 07:28:00 PM FASTING:NO. 2:72MG/DL (B)(6) 2015 07:28:00 PM FASTING:NO. 3:61MG/DL (B)(6) 2015 07:27:00 PM FASTING:NO. 4:60MG/DL (B)(6) 2015 07:25:00 PM FASTING:NO. CUSTOMER IS CONCERN WITH ALL THESE RESULTS BECAUSE THEY WERE TAKEN ON 3 DIFFERENT PERSON AND ALL THE RESULTS ARE LOW. CUSTOMER CALLING STATES THAT HE JUST PURCHASE THE METER AN HOUR AGO AND TESTED 3 DIFFERENT PEOPLE AND DOES NOT TRUST THE METER RESULTS. CHECK MEMORY AND THE TIME ARE NOT SET CORRECTLY. ONLY TWO RESULTS IN THE METER MEMORY BELONG TO THE CUSTOMER RESULTS TAKEN ON (B)(6) 2015 65MG/DL AND 60MG/DL.

Description of Event or Problem · 1

CUSTOMER COMPLAINS OF LOW RESULTS. CUSTOMER STATES THAT HE FEELS WELL AND REQUIRES NO MEDICAL ATTENTION. THE CUSTOMER'S EXPECTED BLOOD RESULTS ARE 130-160MG/DL FASTING. VERIFIED THE STRIPS EXPIRED 3/17/2018. CUSTOMER CONFIRMS THE STRIPS HAVE BEEN PROPERLY STORED AND WERE FIRST OPENED (B)(6) 2015. CUSTOMER PERFORMED A BLOOD TEST AND OBTAINED 64MG/DL. 65MG/DL, (B)(6) 2015 07:28:00 PM, FASTING:NO. 72MG/DL, (B)(6) 2015 07:28:00 PM, FASTING:NO. 61MG/DL, (B)(6) 2015 07:27:00 PM, FASTING:NO.. 4:60MG/DL, (B)(6) 2015 07:25:00 PM, FASTING:NO. CUSTOMER IS CONCERN WITH ALL THESE RESULTS BECAUSE THEY WERE TAKEN ON 3 DIFFERENT PERSON AND ALL THE RESULTS ARE LOW. CUSTOMER CALLING STATES THAT HE JUST PURCHASE THE METER AN HOUR AGO AND TESTED 3 DIFFERENT PEOPLE AND DOES NOT TRUST THE METER RESULTS. CHECK MEMORY AND THE TIME ARE NOT SET CORRECTLY. ONLY TWO RESULTS IN THE METER MEMORY BELONG TO THE CUSTOMER RESULTS TAKEN ON (B)(6) 2015 65MG/DL AND 60MG/DL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
798273 BLOOD GLUCOSE SYSTEM NBW NIPRO DIAGNOSTICS, INC. TRUERESULT PS2246

Patients

Seq Age Sex Outcome Treatment
1 0 YR