FDA Adverse Event Death Summary report: N

EASYCARE BED

MDR report key: 5267874 · Received December 4, 2015

Report

Report Number
3009402404-2015-00036
Event Type
Death
Date Received
December 4, 2015
Date of Event
November 16, 2015
Report Date
December 4, 2015
Manufacturer
JOERNS HEALTHCARE
Product Code
FNL
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NM, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL EQUIPMENT: F14SCAL RAILS ON THE BED, SERIAL#'S (B)(4), MANUFACTURED BY JOERNS HEALTHCARE, DATE OF MANUFACTURER IS 12/08/2011. THE AIR MATTRESS INVOLVED IS MANUFACTURED BY (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED TO THE MANUFACTURER BY THE END USER, PER THE END USER CNA WAS MAKING ROUNDS AND WITNESSED THE RESIDENT HALF WAY ON THE FLOOR WITH HER HEAD POSSIBLE STUCK BETWEEN SIDE RAIL AND AIR MATTRESS. THE PATIENTS' TOP HALF OF HER BODY WAS ON THE BED AND LOWER HALF WAS ON THE FLOOR. BED WAS IN THE LOWEST POSITION. EMS WAS CALLED AND PROVIDED SUPPORTIVE CARE UNTIL A DNR WAS COMPLETED. A BED ASSESSMENT WAS NOT PERFORMED FOR THIS PATIENT PRIOR TO USING THE BED, RAILS OR AIR MATTRESS. ON (B)(6) 015, A JOERNS ASSOCIATE VISITED THE FACILITY AND TESTED THE BED AND RAILS. THE BED AND RAILS ARE OPERATING IN GOOD SHAPE AND ARE IN-SERVICE WITH ANOTHER RESIDENT SINCE INCIDENT. COMPLAINT# (B)(4) WAS ENTERED INTO OUR SYSTEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
797065 EASYCARE BED ELECTRIC PATIENT BED FNL JOERNS HEALTHCARE EASYCARE 5

Patients

Seq Age Sex Outcome Treatment
1 82 YR Death