BIOMEX SHELL PC DIA48MM LN 22
Report
- Report Number
- 3002806535-2015-04113
- Event Type
- Injury
- Date Received
- December 2, 2015
- Date of Event
- August 19, 2015
- Report Date
- January 28, 2016
- Manufacturer
- BIOMET UK LTD.
- Product Code
- LZO
- PMA / PMN Number
- PN/A
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FI
- Reporter Occupation
- PHYSICIAN
Narratives
THIS USER FACILITY IS OUTSIDE OF THE UNITED STATES. CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. THIS PRODUCT IS MANUFACTURED BY ZIMMER BIOMET (B)(4) AND IS NOT CLEARED OR DISTRIBUTED IN THE U.S. HOWEVER, THIS REPORT IS BEING SUBMITTED AS ZIMMER BIOMET IN (B)(4) MANUFACTURES A SIMILAR DEVICE IN THE UNITED STATES UNDER 510K NUMBER K921181. EVALUATION IN PROCESS BUT NOT YET COMPLETE. UPON COMPLETION OF EVALUATION, A FOLLOW UP REPORT WILL BE SENT TO THE FDA.
THIS FOLLOW-UP REPORT IS BEING FILED TO RELAY ADDITIONAL INFORMATION, WHICH WAS UNKNOWN AT THE TIME OF THE INITIAL MEDWATCH. EXAMINATION OF RETURNED DEVICE FOUND NO EVIDENCE OF PRODUCT NON-CONFORMANCE. DURING THE EVALUATION, IT WAS NOTED THE ROOT CAUSE OF THE EVENT WAS MOST LIKELY ATTRIBUTED TO THE DURATION OF PRODUCT BEING IMPLANTED. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THIS TYPE OF EVENT CAN OCCUR: UNDER POSSIBLE ADVERSE EFFECTS, NUMBER 14 STATES, "FATIGUE FRACTURE OF COMPONENT CAN OCCUR AS A RESULT OF LOSS OF FIXATION, STRENUOUS ACTIVITY, MALALIGNMENT, TRAUMA, NON-UNION, OR EXCESSIVE ACTIVITY. NUMBER 16 STATES, "WEAR AND/OR DEFORMATION OF ARTICULATING SURFACES." THIS REPORT IS NUMBER 1 OF 2 MDRS FILED FOR THE SAME PATIENT (REFERENCE 3002806535-2015-04113 AND 0001825034-2015-03913).
IT WAS REPORTED THAT PATIENT UNDERWENT TOTAL HIP ARTHROPLASTY ON (B)(6) 2002. SUBSEQUENTLY, A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2015 DUE TO FRACTURED ACETABULAR LINER AND DAMAGED ACETABULAR SHELL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 792092 | BIOMEX SHELL PC DIA48MM LN 22 | PROSTHESIS, HIP | LZO | BIOMET UK LTD. | N/A | 313077 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |