POS COMBO PANEL TYPE 33
Report
- Report Number
- 2919016-2015-00112
- Event Type
- Malfunction
- Date Received
- December 1, 2015
- Date of Event
- November 2, 2015
- Report Date
- November 2, 2015
- Manufacturer
- BECKMAN COULTER
- Product Code
- JWY
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- 2050012-12/02/2015-009C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE DATA PROVIDED BY THE CUSTOMER WAS REVIEWED AND CONFIRMED THE HIGH PROBABILITY MISIDENTIFICATION OF S. AUREUS AS S. INTERMEDIUS. THERE WAS NO INDICATION OF INSTRUMENT ISSUES AND RAW DATA AND PROCESSED VALUES FOR CONTROL AND GROWTH WELLS WERE SATISFACTORY. PYR VALUES WERE CONFIRMED AS POSITIVE ON THE PROCESSED PANEL REPORTS. POS COMBO 33 PANEL, LOT 2016-05-04 IS SUBJECT TO A BECKMAN COULTER FIELD ACTION DUE TO FALSELY NEGATIVE RESULTS FOR VOGES-PROSKAUER WITH QUALITY CONTROL AMERICAN TYPE CULTURE COLLECTION ORGANISM STAPHYLOCOCCUS AUREUS ATCC 29213. (B)(4).
BECKMAN COULTER ISSUED A FIELD ACTION FA-26321 AND AN URGENT MEDICAL DEVICE RECALL LETTER WAS SENT VIA (B)(4) TO THE AFFECTED CUSTOMERS ON 12/04/2015. THE URGENT MEDICAL DEVICE RECALL LETTER INFORMS CUSTOMERS THAT FALSELY NEGATIVE RESULT WAS TRACED TO A SINGLE VOGES-PROSKAUER (VP) COMPONENT LOT USED IN THE MANUFACTURING OF THE MICROSCAN POS COMBO PANEL TYPE 33 LOT 2016-05-04. THE CUSTOMERS WERE ADVISED TO DISCONTINUE USE AND DISCARD ANY REMAINING INVENTORY OF THE AFFECTED LOT. THE VP COMPONENT WHICH RESULTED IN OUT OF SPECIFICATION RESULTS WITH QUALITY CONTROL STAPHYLOCOCCUS AUREUS ATCC 29213 IS NO LONGER IN USE. (B)(4).
IT WAS REPORTED THAT POS COMBO 33 PANEL, LOT 2016-05-04 RESULTED TO HIGH PROBABILITY MISIDENTIFICATION OF STAPHYLOCOCCUS AUREUS AS S. INTERMEDIUS. THE PYRROLIDONYL-B-NAPHTHYLAMIDE (PYR) RESULTED IN POSITIVE, HOWEVER THE PYR PANEL WELL REACTION SHOULD HAVE BEEN NEGATIVE, ACCORDING TO THE CUSTOMER'S ALTERNATE METHOD . VISUAL VERIFICATION OF THE PYR PANEL WELL CONFIRMED THE POSITIVE RESULT. THE ISOLATE WAS NOT SENT TO A REFERENCE LABORATORY FOR CONFIRMATORY TESTING, RATHER THE REPORTED ISOLATE MISIDENTIFICATION WAS BASED ON THE CUSTOMER'S ALTERNATE METHOD I.E. LATEX AGGLUTINATION TEST. THERE WERE NO ERRONEOUS RESULTS REPORTED OUT OF THE LABORATORY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 787418 | POS COMBO PANEL TYPE 33 | MICRO DILUTION PANEL | JWY | BECKMAN COULTER | N/A | 2016-05-04 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |