PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 5253322 · Received November 28, 2015

Report

Report Number
2032227-2015-69387
Event Type
Malfunction
Date Received
November 28, 2015
Date of Event
November 9, 2015
Report Date
November 9, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH NO A44 ALARM NOTED AND PASSED SELF TEST. HOWEVER, THE INSULIN PUMP HAD MINOR SCRATCHED LCD WINDOW, CRACKED CASE NEAR THE DISPLAY WINDOW CORNERS, BROKEN BELT CLIP SLOT, CRACKED RESERVOIR TUBE LIP AND CRACKED RESERVOIR TUBE.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED VIA PHONE CALL MULTIPLE IOP TEST FAILURE ALARMS ON THE INSULIN PUMP. CUSTOMER ADVISED THEY RECEIVED THE ALARM TWICE DURING A BASAL ATTEMPT. TROUBLESHOOTING FOR THE ALARM WAS PERFORMED. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND REVERT TO A BACKUP PLAN. CUSTOMER WAS ADVISED THAT THE DEVICE WOULD BE REPLACED AND AGREED TO RETURN THE PRODUCT FOR ANALYSIS.