FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 5252946 · Received November 27, 2015

Report

Report Number
2531779-2015-43077
Event Type
Malfunction
Date Received
November 27, 2015
Report Date
November 17, 2015
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 01/12/2106 DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 12/30/2015 WITH THE FOLLOWING FINDINGS: THE ALLEGED ISSUE OF A DUAL ALARM FAILURE COULD NOT BE DUPLICATED. UNRELATED TO THE INITIAL ALLEGATION, THE BATTERY COMPARTMENT WAS CRACKED.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2015, THE REPORTER CONTACTED ANIMAS, ALLEGING A AUDIO TONE/VIBRATION (DUAL ALARM FAILURE) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
782863 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 71 YR