DIMENSION® SIROLIMUS FLEX® REAGENT CARTRIDGE
Report
- Report Number
- 2517506-2015-00210
- Event Type
- Malfunction
- Date Received
- November 26, 2015
- Date of Event
- November 1, 2015
- Report Date
- February 2, 2016
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
- Product Code
- NRP
- PMA / PMN Number
- K081857
- Removal / Correction Number
- 2517506-02/12/2016-001-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
THE CAUSE OF THE IMPRECISION WITH QC AND PATIENT SAMPLES WITH SIROLIMUS LOT EB6064 IS UNKNOWN. THE CUSTOMER MOVED TESTING TO AN ALTERNATE LOT. THE ISSUE IS UNDER INVESTIGATION BY SIEMENS HEALTHCARE DIAGNOSTICS.
SIEMENS HEALTHCARE DIAGNOSTICS RECEIVED CUSTOMER COMPLAINTS FOR DIMENSION SIROLIMUS (SIRO) LOT EB6064. THE COMPLAINTS WERE RELATED TO QC SHIFTS AND IMPRECISION. SIEMENS HEALTHCARE DIAGNOSTICS CONDUCTED AN INVESTIGATION ON THE ON BOARD STABILITY FOR DIMENSION SIRO LOT EB6064. INVESTIGATION DETERMINED THAT DIMENSION SIRO LOT EB6064 DOES NOT MEET THE OPEN WELL STABILITY CLAIM OF TWO DAYS. INTERNAL TESTING HAS SHOWN BIASES RANGING FROM -82% TO +123% ON A POOLED SAMPLE WITH A SIROLIMUS CONCENTRATION O 8.7 NG/ML [9.5 MMOL/L] WHEN TESTED ON SUBSEQUENT DAYS OF THE TWO DAY OPEN WELL STABILITY PERIOD. RESULTS ARE UNAFFECTED AT THE START OF THE OPEN WELL STABILITY PERIOD. QUALITY CONTROL IS LIKELY TO CATCH THE ISSUE ON SUBSEQUENT DAYS OF THE OPEN WELL STABILITY CLAIM. SIEMENS HEALTHCARE DIAGNOSTICS ISSUED AN URGENT MEDICAL DEVICE RECALL COMMUNICATION DC-16-01.B.US DATED FEBRUARY, 2016 INSTRUCTING CUSTOMERS TO DISCONTINUE USE AND DISCARD ANY REMAINING INVENTORY OF THE AFFECTED LOT AND TO CONTACT SIEMENS HEALTHCARE FOR REPLACEMENT PRODUCT.
A CUSTOMER REPORTED IMPRECISION OF SIROLIMUS (SIRO) RESULTS WITH QC AND PATIENT SAMPLES WITH LOT EB6064. IT IS UNKNOWN IF THE INITIAL PATIENT RESULTS HAVE BEEN REPORTED TO PHYSICIANS. THE SAMPLES WERE REPEATED ON AN ALTERNATE LOT OF REAGENT AND THE REPEAT RESULTS WITH LOT FA6136 WERE REPORTED. IT IS NOT ALLEGED THAT PATIENT TREATMENT WAS ALTERED OR PRESCRIBED ON THE BASIS OF THE IMPRECISE SIRO RESULTS. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE IMPRECISE SIRO RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 782373 | DIMENSION® SIROLIMUS FLEX® REAGENT CARTRIDGE | DIMENSION® SIROLIMUS FLEX® REAGENT CARTRIDGE | NRP | SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW | EB6064 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |