FDA Adverse Event Malfunction Summary report: N

NAVLOCK TRACKER UNIVERSAL ORANGE

MDR report key: 5252009 · Received November 25, 2015

Report

Report Number
1723170-2015-01462
Event Type
Malfunction
Date Received
November 25, 2015
Date of Event
October 30, 2015
Report Date
June 2, 2016
Manufacturer
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
Product Code
OLO
PMA / PMN Number
K124004
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

CHANGING TO NAVLOCK TRACKER AS THE TRACKER IS THE INSTRUMENT THAT WAS REPLACED. THE INTERFACE BETWEEN THE TAP AND THE TRACKER CAUSED THE ISSUE.

Additional Manufacturer Narrative · 1

DEVICE MANUFACTURING DATE IS DEPENDENT ON LOT NUMBER, THEREFORE, UNAVAILABLE. RETURN REQUESTED. REPLACEMENT SOLERA AWL TIP TAP 4.5MM TAP SHIPPED TO SITE (B)(6) 2015. NO PARTS HAVE BEEN RECEIVED BY MANUFACTURER FOR ANALYSIS. NO FURTHER ISSUES HAVE BEEN REPORTED.

Additional Manufacturer Narrative · 1

CORRECTION: BRAND NAME & CATALOG NUMBER. DEVICE LOT NUMBER NOW PROVIDED. DEVICE MANUFACTURING DATE NOW PROVIDED.

Additional Manufacturer Narrative · 1

ALTHOUGH THE COMPLAINT ALLEGES A SINGLE TRACKER WAS DAMAGED, 3 TOTAL TRACKERS WERE RETURNED FOR ANALYSIS: LOT# 100604 - MANUFACTURER DUE: 06/04/2010 AND 2 OF LOT # 150817 - MANUFACTURE DATE: 08/17/2015. MEDTRONIC INVESTIGATION OF RETURNED SUSPECT TRACKER FINDS THAT THE REPORTED ISSUE COULD NOT BE DUPLICATED. THE RETURNED TRACKERS WERE FOUND TO BE IN GOOD CONDITION WITH NO APPARENT PHYSICAL DAMAGE. KNOWN GOOD INSTRUMENTS INSERTED AND ROTATED WITHOUT ISSUE. WITH MARKERS ATTACHED AND FULLY, THE TRACKERS RETURNED GOOD GEOMETRY AND DIVOT ERROR READINGS. NO PROBLEM FOUND.

Description of Event or Problem · 1

A MEDTRONIC REPRESENTATIVE REPORTED THAT, WHILE IN A SPINE PROCEDURE, THE SITE'S SOLERA 4.5 AWL TIP TAP WAS DAMAGED. WHILE USING THE (B)(4) SYSTEM, THE TAP WAS WORKING SLOWLY THEN STOPPED SPINNING. THE TAP APPEARED TO BE STUCK ON THE TRACKER. THE SURGEON OPTED TO PROCEED WITH THE PROCEDURE WITH MANUAL TAPPING. DELAY IN THERAPY WAS LESS THAN ONE HOUR. THE SURGEON COMPLETED THE PROCEDURE WITH THE USE OF THE NAVIGATION SYSTEM. THERE WAS NO IMPACT ON PATIENT OUTCOME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
779218 NAVLOCK TRACKER UNIVERSAL ORANGE ORTHOPEDIC STEREOTAXIC INSTRUMENT OLO MEDTRONIC NAVIGATION, INC. (LOUISVILLE) 150817

Patients

Seq Age Sex Outcome Treatment
1 67 YR