ALARIS® PUMP MODULE ADMINISTRATION SET
Report
- Report Number
- 9616066-2015-01684
- Event Type
- Malfunction
- Date Received
- November 25, 2015
- Report Date
- November 11, 2015
- Manufacturer
- CAREFUSION
- Product Code
- FPA
- PMA / PMN Number
- K944320
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
ALTHOUGH REQUESTED, THE AFFECTED PRODUCT HAS NOT BEEN RECEIVED. A FOLLOW UP REPORT WILL BE SUBMITTED WITH INVESTIGATION RESULTS SHOULD THE DEVICES BE RECEIVED FOR EVALUATION.
CONCOMITANT MEDICAL PRODUCTS: BAXTER, 500ML IV BAG OF 0.9% SODIUM CHLORIDE, LOT: C990002, EXP: DECEMBER 2016. THE CUSTOMER¿S REPORT OF IV TUBING SEPARATION ¿TEAR¿ WAS CONFIRMED. VISUAL INSPECTION OBSERVED THAT THE SILICONE SEGMENT HAD A TEAR NEAR THE UPPER FITMENT. THE TEAR MEASURED 0.0581 INCHES LONG. VISUAL EXAMINATION UNDER A MICROSCOPE NOTED NO CRUSH MARKS TO THE UPPER BLUE FITMENT. NO OTHER ANOMALIES WERE NOTED DURING VISUAL INSPECTION. FUNCTIONAL TESTING WAS PERFORMED; A LEAK WAS OBSERVED COMING OUT FROM THE SILICONE TUBING NEAR THE UPPER FITMENT. THE SET WAS PRESSURE TESTED WHILE SUBMERGED UNDERWATER AT 30 PSI OF AIR; LEAKING WAS OBSERVED AT THE SILICONE TUBING AT LESS THAN 3PSI OF AIR. THE ROOT CAUSE OF THE TEAR WAS NOT IDENTIFIED.
THE CUSTOMER REPORTED A SEPARATION OF AN IV TUBING DURING AN INFUSION OF AN UNSPECIFIED MEDICATION OR FLUID. THERE IS NO REPORT OF PATIENT HARM; ADDITIONAL DETAILS ARE NOT AVAILABLE AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 781127 | ALARIS® PUMP MODULE ADMINISTRATION SET | SET, ADMINISTRATION, INTRAVASCULAR | FPA | CAREFUSION | 2420-0500 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |