DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM, DEXCOM SHARE SYSTEM
Report
- Report Number
- 3004753838-2015-77578
- Event Type
- Malfunction
- Date Received
- November 25, 2015
- Date of Event
- November 5, 2015
- Report Date
- November 5, 2015
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER
Narratives
(B)(4).
(B)(4).
PATIENT'S MOTHER CONTACTED DEXCOM ON (B)(6) 2015 TO REPORT THAT ON (B)(6) 2015, PATIENT EXPERIENCED AN INTERMITTENT OUT OF RANGE SIGNAL. THE DEVICE HAS BEEN RECEIVED FOR EVALUATION. A FOLLOW UP REPORT WILL BE SUBMITTED ONCE THE EVALUATION HAS BEEN COMPLETED. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE.
SUBSEQUENT TO THE INITIAL MDR, THE COMPLAINT RECEIVER WAS NOT RETURNED TO DEXCOM. THE TRANSMITTER (PART NUMBER STT-GL-003/SERIAL NUMBER (B)(4)/LOT NUMBER 5201938), BEING USED WITH THE COMPLAINT RECEIVER, WAS RETURNED ON (B)(6) 2015. THE RETURNED TRANSMITTER WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE TESTS FAILED. THE REPORTED EVENT WAS CONFIRMED. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE TRANSMITTER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 779892 | DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM, DEXCOM SHARE SYSTEM | MDS | MDS | DEXCOM, INC. | MT22608-BLU | 5201938 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 4 YR |