FDA Adverse Event Malfunction Summary report: N

NEOPUFF INFANT RESUSCITATOR

MDR report key: 5242009 · Received November 22, 2015

Report

Report Number
9611451-2015-00492
Event Type
Malfunction
Date Received
November 22, 2015
Report Date
October 26, 2015
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BTL
PMA / PMN Number
K892885
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT NEOPUFF WAS RECEIVED AT OUR REGIONAL OFFICE IN (B)(6), WHERE IT WAS INSPECTED BY A TRAINED FISHER & PAYKEL HEALTHCARE SERVICE ENGINEER. THE FASCIA AND VALVE SYSTEM WERE REPLACED AND THE NEOPUFF WAS RETURNED TO THE CUSTOMER AFTER A PERFORMANCE TEST. A PHOTOGRAPH OF THE SUBJECT NEOPUFF SHOWING THE DAMAGE WERE ALSO PROVIDED. RESULTS: VISUAL INSPECTION OF THE PROVIDED PHOTOGRAPH REVEALED PHYSICAL DAMAGE TO THE GAS OUTLET PORT THAT APPEARED TO BE THE RESULT OF IMPACT TO THE DEVICE. A LOT CHECK REVEALED ONE OTHER COMPLAINT FOR LOT 070731. CONCLUSION: THE NEOPUFF IS A PORTABLE, REUSABLE DEVICE USED TO ASSIST IN THE DELIVERY OF RESPIRATORY BREATHS TO INFANTS UNTIL ADEQUATE SPONTANEOUS BREATHING OCCURS. BEING A PORTABLE DEVICE, THE NEOPUFF CAN BE SUSCEPTIBLE TO IMPACT DAMAGE, FOR INSTANCE WHEN ACCIDENTALLY DROPPED OR SUBJECTED TO CONSIDERABLE EXTERNAL FORCE. THE NEOPUFF TECHNICAL MANUAL WARNS AGAINST DROPPING THE NEOPUFF OR SUBJECTING IT TO IMPACT DAMAGE WHICH MAY CAUSE THE UNIT TO OPERATE INCORRECTLY. IF THE NEOPUFF IS SUSPECTED TO HAVE BEEN DAMAGED, THE MANOMETER AND VALVE SYSTEM SHOULD BE PERFORMANCE TESTED. THE NEOPUFF TECHNICAL MANUAL STATES THE FOLLOWING: DROPPING THE NEOPUFF / PERIVENT INFANT RESUSCITATOR OR OTHER SIMILAR FORMS OF IMPACT MAY CAUSE DAMAGE RESULTING IN INCORRECT OPERATION OF THE UNIT. IF YOU SUSPECT DAMAGE TO HAVE OCCURRED, PLEASE PERFORM CHECKS AS OUTLINED IN THE NEOPUFF TECHNICAL MANUAL BEFORE CONNECTION TO A PATIENT.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED THAT A NEOPUFF INFANT RESUSCITATOR GAS OUTLET VALVE WAS BROKEN DURING USE AND REQUESTED SERVICE OF THE DEVICE. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
772249 NEOPUFF INFANT RESUSCITATOR BTL BTL FISHER & PAYKEL HEALTHCARE LTD RD900 070731

Patients

Seq Age Sex Outcome Treatment
1