PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2015-68095
- Event Type
- Malfunction
- Date Received
- November 20, 2015
- Date of Event
- October 29, 2015
- Report Date
- October 29, 2015
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN
- Reporter Occupation
- PATIENT
Narratives
THE INSULIN PUMP HAD INTERMITTENT ESCAPE AND ACT BUTTON RESPONSE DUE TO FLATTENED DOME SWITCHES. NO BUTTON ERROR ALARM WAS NOTED DURING TESTING. THE KEYPAD CONNECTOR WAS LOCKED. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON THE DISPLAY WINDOW, SCRATCHED RESERVOIR TUBE WINDOW AND A BROKEN BELT CLIP SLOT. (B)(4).
THE CUSTOMER REPORTED VIA TELEPHONE CALL THAT THE INSULIN PUMP ALARMED FOR A BUTTON ERROR. THE BLOOD GLUCOSE READING WAS 94 MG/DL. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO A BACK UP PLAN PER A HEALTH CARE PROFESSIONAL'S INSTRUCTIONS. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP WOULD BE REPLACED AND AGREED TO RETURN THE PRODUCT FOR ANALYSIS.