FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 5240986 · Received November 20, 2015

Report

Report Number
1720753-2015-04716
Event Type
Malfunction
Date Received
November 20, 2015
Date of Event
November 4, 2015
Report Date
November 20, 2015
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
OWB
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE HARD DISK DRIVE AND USB HUB WERE EVALUATED AND REPLACED. THE SYSTEM SOFTWARE AND CALIBRATIONS WERE ALSO RELOADED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM INTERMITTENTLY FAILED TO BOOT AND EXHIBITED INTERMITTENT SYSTEM FUNCTIONALITY. NO PATIENT SERIOUS INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
770434 9900 FLUOROSCOPIC X-RAY OWB GE OEC MEDICAL SYSTEMS (SLC) 9900 E9-0425

Patients

Seq Age Sex Outcome Treatment
1