FDA Adverse Event Injury Summary report: N

ERBE VIO 300D

MDR report key: 5239765 · Received November 20, 2015

Report

Report Number
9610614-2015-00015
Event Type
Injury
Date Received
November 20, 2015
Date of Event
August 27, 2015
Report Date
November 20, 2015
Manufacturer
ERBE ELEKTROMEDIZIN GMBH
Product Code
GEI
PMA / PMN Number
K083452
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE MEDICAL FACILITY REPORTED THAT USER ERROR CAUSED THE REPORTED ISSUE AND THEY DO NOT CONSIDER THE INCIDENT TO BE RELATED TO ERBE EQUIPMENT. NO ANOMALIES WERE FOUND UPON THE REVIEW OF THE UNIT'S DEVICE HISTORY RECORD (DHR). TO FURTHER ADDRESS THE ISSUE, ADDITIONAL IN-SERVICE TRAINING WAS OFFERED. THE ACCOUNT IS BEING MADE AWARE OF THE FINDINGS. ERBE USA, INC. IS NOW CLOSING THE FILE ON THIS EVENT.

Description of Event or Problem · 1

THE ACCOUNT VOLUNTARILY REPORTED TO FDA IN REPORT NUMBER MW5057523 THE FOLLOWING: "THE PT HAD A HYSTEROSCOPIC REMOVAL OF POLYP AND D&C ON (B)(6) 2015. THE OPERATING HYSTEROSCOPE WAS PLACED WITH A LOOP CAUTERY. ATTEMPT WAS MADE TO GET TO THE BASE BUT BECAUSE OF ANGULATION, THE CAUTERY COULD NOT BE UTILIZED. THE HYSTEROSCOPE WAS REMOVED AND THE POLYP FORCEPS WERE USED TO REMOVE THE POLYP. A SHARP CURETTAGE WAS PERFORMED. HYSTEROSCOPE WAS AGAIN PLACED AND THE BASE WAS CAUTERIZED USING CUTTING CURRENT. IT WAS NOTED THAT THE SODIUM CHLORIDE LOSS WAS UP TO 700 ML. THE UTERINE CAVITY WAS VISUALIZED WITH NO OBVIOUS PERFORATION VISIBLE. ON (B)(6) 2015, THE PT PRESENTED TO THE ED FOR THE ABRUPT ONSET OF ABDOMINAL PAIN. THE PT WAS BROUGHT TO THE OPERATING ROOM FOR AN EXPLORATORY LAPAROTOMY FOR PERFORATION OF THE UTERUS S/P HYSTEROSCOPY AND SMALL BOWEL INJURY. THE SMALL INTESTINE WAS NOTED TO HAVE A FOCAL PERFORATION WHICH WAS THOUGHT TO BE A THERMAL INJURY, WHICH WAS CLOSED WITH SUTURES."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
769634 ERBE VIO 300D ELECTROSURGICAL UNIT GEI ERBE ELEKTROMEDIZIN GMBH VIO300D

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R