FDA Adverse Event Injury Summary report: N

QUADRA H CEMENTLESS STEM SIZE 5 STD

MDR report key: 5239251 · Received November 20, 2015

Report

Report Number
3005180920-2015-00281
Event Type
Injury
Date Received
November 20, 2015
Date of Event
October 22, 2015
Report Date
February 18, 2016
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JDI
PMA / PMN Number
K082792
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ON 20 JANUARY 2016, IT WAS PREPARED A FINAL REPORT WITH THE INFORMATION ALREADY SUBMITTED IN THE INTIAL REPORT. ON 25 JANUARY 2016, THE REPORT WAS SENT TO THE INITIAL REPORTER AND THE CASE WAS CLOSED.

Additional Manufacturer Narrative · 1

ON 12 NOV 2015 IT WAS COMMUNICATED THAT REV. REASON: THIGH PAIN, VISIBLE DUE TO LOOSENING OF THE STEM ON RX. REV. ROUTINE: STEM AND HEAD CHANGED. BATCH REVIEW PERFORMED ON 17 NOVEMBER 2015: LOT 113027: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 13 DECEMBER 2011. EXPIRATION DATE: 2016-09-30. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, ALL THE ITEMS OF THE LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. ON 18 NOV 2015 THE MEDICAL AFFAIRS DIRECTOR MADE THE FOLLOWING ANALYSIS: FEMORAL STEM REVISION 3,5 YEARS AFTER PRIMARY IMPLANT DUE TO FEMORAL ASEPTIC LOOSENING. FROM THE XRAY, LOOSENING IS VISIBLE THOUGH NOT VERY EVIDENT, BUT I CANNOT DISCERN ANY PLAUSIBLE CAUSE FOR THIS OCCURRENCE THAT CAN BE RELATED TO A CLINICAL MISTAKE. APPARENTLY, LOOSENING STARTED IN THE PROXIMAL REGION AND IT IS MORE EVIDENT PROXIMALLY. A MALE PATIENT, WITH DORR TYPE A FEMUR, STRONG AND THICK CORTICAL BONE IN THE CANAL IS POSSIBLY MORE LIKELY TO GET THIS KIND OF COMPLICATION, WHICH IS RARE BUT DESCRIBED IN LITERATURE. ON 19 NOV 2015 IT WAS CONFIRMED THAT THE ITEM WILL NOT BE AVAILABLE FOR ANALYSIS.

Description of Event or Problem · 1

REVISION BECAUSE OF PAINFULLY STEM LOOSENING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
769646 QUADRA H CEMENTLESS STEM SIZE 5 STD 01.12.025 JDI MEDACTA INTERNATIONAL SA 113027

Patients

Seq Age Sex Outcome Treatment
1 68 YR Required Intervention