ACCENT DR RF
Report
- Report Number
- 2017865-2015-30322
- Event Type
- Malfunction
- Date Received
- November 19, 2015
- Report Date
- October 5, 2018
- Manufacturer
- ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
- Product Code
- LWP
- PMA / PMN Number
- P880086
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4).
ANALYSIS WAS NORMAL, NO ANOMALIES WERE NOTED.
CORRECTION: THE PMA/510(K) # SHOULD HAVE BEEN P880086 RATHER THAN P960013.
IT WAS REPORTED THAT THE PATIENT EXPERIENCED DIZZY SPELLS. DURING FOLLOW UP, THE PULSE GENERATOR EXHIBITED VENTRICULAR NOISE. ELECTROGRAMS REVEALED PACING INHIBITION DURING NOISE. THE NOISE COULD BE REPRODUCED WITH POCKET MANIPULATION. THE DEVICE WAS REPROGRAMMED. THE PATIENT WOULD BE MONITORED.
NEW INFORMATION INDICATED THAT THE PULSE GENERATOR CONTINUED TO EXHIBIT NOISE, RESULTING IN PACING INHIBITION. THE PATIENT CONTINUED TO EXPERIENCE LIGHTHEADEDNESS. THE NOISE COULD NOT BE REPRODUCED.
NEW INFORMATION INDICATED THAT THE PULSE GENERATOR WAS EXPLANTED AND REPLACED ON (B)(6) 2016 DURING RIGHT VENTRICULAR LEAD REVISION PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES AND THE PATIENT WAS WELL POST PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 766342 | ACCENT DR RF | IMPLANTABLE PACEMAKER PULSE GENERATOR | LWP | ST. JUDE MEDICAL, INC.(CRM-SYLMAR) | PM2212 | 3204839 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |