MDR report key: 5233813 · Received November 18, 2015

Report

Report Number
1220908-2015-03010
Event Type
Malfunction
Date Received
November 18, 2015
Report Date
November 2, 2015
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
PMA / PMN Number
K060559
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION; THE MALFUNCTION WAS DUPLICATED AND ATTRIBUTED TO A DAMAGED ETCH ON THE PACE/DEFIB ENGINE BOARD. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED "DEFIB DISABLED" AND "PACER DISABLED" MESSAGES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
762470 R SERIES DEFIBRILLATOR DEFIBRILLATOR/PACEMAKER MKJ ZOLL MEDICAL CORPORATION R SERIES NA

Patients

Seq Age Sex Outcome Treatment
1