FDA Adverse Event Malfunction Summary report: N

POWER CHAIR 20 WIDE BY 18 DEEP FOR PMI

MDR report key: 5233712 · Received November 18, 2015

Report

Report Number
3008262382-2015-01957
Event Type
Malfunction
Date Received
November 18, 2015
Report Date
November 18, 2015
Manufacturer
INVACARE REHABILITATION EQUIPMENT CO.
Product Code
ITI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

PRODUCT WAS RETURNED FOR EVALUATION. THE RETURN FIELDS IN ORACLE STATE: CONSUMER POWER WHEELCHAIRS. MOTOR, BRAKE FAILURE - ELECTRICAL. ELECTRICAL BRAKE FAILURE - BENCH TEST. WIGGLE WIRES BY THE STRAIN RELIEF AND IT WILL SHUT OFF AND GIVE A BRAKE FAULT DURING BENCH TEST. PRODUCT RECEIVED EXPANDED EVALUATION. THE EXPANDED EVALUATION REPORT STATES: UTILIZING EXISTING COMPLAINT INFORMATION, ACTUAL OBSERVATIONS, AND FUNCTIONAL TESTING OF THE RETURNED PRODUCT IN ITS "AS RECEIVED" CONDITION, THE COMPLAINT WAS CONFIRMED FOR THE LEFT MOTOR AND GEARBOX OF HAVING A BRAKE FAULT CAUSED BY A BAD CONNECTION NEAR THE STRAIN RELIEF. COMPLAINT WAS CONFIRMED. THE UNDERLYING CAUSE IS BAD CONNECTION NEAR THE STRAIN RELIEF.

Additional Manufacturer Narrative · 1

SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL RECORD WILL BE FILED.

Description of Event or Problem · 1

PRODUCT WAS RETURNED FOR EVALUATION. THE RETURN FIELDS IN ORACLE STATE: CONSUMER POWER WHEELCHAIRS. MOTOR, BRAKE FAILURE - ELECTRICAL. ELECTRICAL BRAKE FAILURE - BENCH TEST. WIGGLE WIRES BY THE STRAIN RELIEF AND IT WILL SHUT OFF AND GIVE A BRAKE FAULT DURING BENCH TEST. PRODUCT RECEIVED EXPANDED EVALUATION. THE EXPANDED EVALUATION REPORT STATES: UTILIZING EXISTING COMPLAINT INFORMATION, ACTUAL OBSERVATIONS, AND FUNCTIONAL TESTING OF THE RETURNED PRODUCT IN ITS "AS RECEIVED" CONDITION, THE COMPLAINT WAS CONFIRMED FOR THE LEFT MOTOR AND GEARBOX OF HAVING A BRAKE FAULT CAUSED BY A BAD CONNECTION NEAR THE STRAIN RELIEF. DEALER IS STATING THAT THE MOTOR LEAD IS LOOSE, AND EVERYTIME IT MOVES THE MOTOR SHUTS DOWN.

Description of Event or Problem · 1

DEALER IS STATING THAT THE MOTOR LEAD IS LOOSE, AND EVERYTIME IT MOVES, THE MOTOR SHUTS DOWN.