BONE PIN, 3.2 X 80 MM, STERILE (2 PACK)
Report
- Report Number
- 3005985723-2015-00252
- Event Type
- Malfunction
- Date Received
- November 17, 2015
- Date of Event
- November 11, 2015
- Report Date
- November 11, 2015
- Manufacturer
- MAKO SURGICAL CORP.
- Product Code
- OLO
- PMA / PMN Number
- K112507
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.
REPORTED EVENT: DURING PARTIAL KNEE REPLACEMENT SURGERY, THE TIP OF A 3.2X80MM BONE PIN BROKE OFF AND EMBEDDED IN PATIENT. THE FRAGMENT OF BONE PIN WAS LEFT IN THE PATIENT'S BONE AND CONFIRMED IN POST-OP X-RAY. DEVICE EVALUATION AND RESULTS: VISUAL INSPECTION WAS NOT PERFORMED AS THE DEVICE WAS NOT RETURNED FOR EVALUATION. DIMENSION INSPECTION COMPLETED AT THE TIME OF MANUFACTURING SHOWS THE LOT WAS WITHIN SPECIFICATION. DEVICE HISTORY REVIEW: REVIEW OF THE DEVICE HISTORY RECORDS FOR THE ASSOCIATED LOT INDICATED (B)(4) DEVICES (111671 NON-STERILE BONE PIN, SINGLE) WERE MANUFACTURED SHIPPED TO (B)(4) ON 08/21/2015 AND ACCEPTED INTO FINAL STOCK ON 08/21/2015 PER ERP. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS IN TRACKWISE AND CATSWEB RELATED TO P/N 111670, LOT NUMBER W39919 SHOWS NO COMPLAINTS RELATED TO THE FAILURE IN THIS INVESTIGATION. TRACKING OF COMPLAINTS RELATED TO THE 111670 PART NUMBER WILL BE TRACKED THROUGH QUARTERLY TREND REQUEST # (B)(4). CONCLUSIONS: AS PART OF THE INVESTIGATION INTO THIS COMPLAINT, THE STRYKER REP THAT WAS PRESENT FOR THE CASE WAS CONTACTED ON THREE SEPARATE OCCASIONS TO DETERMINE IF THE PARALLEL DRILL GUIDE WAS USED. AFTER THESE ATTEMPTS TO DETERMINE IF THE DRILL GUIDE WAS USED DURING PLACEMENT OF THE BONE PINS (FIRST REQUEST WAS (B)(6) 2015; FINAL ATTEMPT WAS (B)(6) 2016) NO RESPONSE WAS RECEIVED REGARDING THE USE OF THE DRILL GUIDE. IT CANNOT BE DETERMINED IF THE SURGEON USED THE DRILL GUIDE DURING PLACEMENT OF THE BONE PINS. HOWEVER, TO DATE FOUR COMPLAINT INVESTIGATIONS HAVE BEEN COMPLETED FOR THIS PART NUMBER. OF THOSE FOUR COMPLAINT INVESTIGATIONS, NO COMPLAINTS HAVE BEEN INVESTIGATED WHERE THE BONE PINS WERE INSERTED USING A DRILL GUIDE AND WERE BROKEN INTRAOPERATIVELY . IT IS POSSIBLE THAT THE SURGEON DID NOT USE THE DRILL GUIDE DURING THE PLACEMENT OF THE BONE PINS AS THE DEVICES WERE WITHIN SPECIFICATION ACCORDING TO THE DHR DOCUMENTATION INCLUDED IN ATTACHMENT 1. CORRECTIVE ACTION/PREVENTIVE ACTION: NO ACTION IS REQUIRED AT THIS TIME AS THERE IS NO INDICATION TO SUGGEST A PRODUCT NON-CONFORMITY OR UNANTICIPATED HAZARD.
THE SURGEON PERFORMED A PARTIAL KNEE ARTHROPLASTY PROCEDURE USING THE ROBOTIC ARM INTERACTIVE ORTHOPEDIC SYSTEM (RIO). DURING THE CASE, THE TIP OF THE BONE PIN BROKE OFF IN THE PATIENT'S TIBIA. IT WAS DISCOVERED ON POST-OP X-RAY.THE SURGEON DECIDED TO LEAVE THE TIP OF THE BONE PIN IN THE PATIENT'S TIBIA. THE OUTCOME OF THE CASE WAS SUCCESSFUL.
THE SURGEON PERFORMED A PARTIAL KNEE ARTHROPLASTY PROCEDURE USING THE ROBOTIC ARM INTERACTIVE ORTHOPEDIC SYSTEM (RIO). DURING THE CASE, THE TIP OF THE BONE PIN BROKE OFF IN THE PATIENT'S TIBIA. IT WAS DISCOVERED ON POST-OP X-RAY. THE SURGEON DECIDED TO LEAVE THE TIP OF THE BONE PIN IN THE PATIENT'S TIBIA. THE OUTCOME OF THE CASE WAS SUCCESSFUL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 760394 | BONE PIN, 3.2 X 80 MM, STERILE (2 PACK) | STEREOTACTIC DEVICE, ACCESSORY | OLO | MAKO SURGICAL CORP. | W39919-1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR | Other |