FDA Adverse Event Malfunction Summary report: N

DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM

MDR report key: 5230297 · Received November 17, 2015

Report

Report Number
3004753838-2015-66626
Event Type
Malfunction
Date Received
November 17, 2015
Date of Event
October 24, 2015
Report Date
October 24, 2015
Manufacturer
DEXCOM, INC.
Product Code
MDS
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ME, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

SUBSEQUENT TO THE INITIAL MDR, THE COMPLAINT RECEIVER WAS NOT RETURNED TO DEXCOM. THE TRANSMITTER (PART NUMBER STT-GL-003/SERIAL NUMBER (B)(4)LOT NUMBER 5201199), BEING USED WITH THE COMPLAINT RECEIVER, WAS RETURNED ON (B)(6) 2015. THE RETURNED TRANSMITTER WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE TEST FAILED. THE REPORTED EVENT OF AN UNRECOVERABLE LOSS OF ANTENNA WAS CONFIRMED. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE TRANSMITTER.

Description of Event or Problem · 1

PATIENT CONTACTED DEXCOM ON (B)(6) 2015, TO REPORT THAT ON (B)(6) 2015 THEIR RECEIVER HAD UNRECOVERABLE LOSS OF ANTENNA. THERE WAS NO REPORT OF INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
759932 DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM MDS MDS DEXCOM, INC. MT22495 5204943

Patients

Seq Age Sex Outcome Treatment
1 64 YR