DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM
Report
- Report Number
- 3004753838-2015-66613
- Event Type
- Malfunction
- Date Received
- November 17, 2015
- Date of Event
- October 19, 2015
- Report Date
- October 23, 2015
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(6).
(B)(4).
PATIENT'S MOTHER CONTACTED DEXCOM ON (B)(6) 2015, TO REPORT THAT ON (B)(6) 2015 THE PATIENT'S RECEIVER HAD UNRECOVERABLE LOSS OF ANTENNA. THERE WAS NO REPORT OF INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. THE DEVICE HAS BEEN RECEIVED FOR EVALUATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE THE EVALUATION IS COMPLETE.
PATIENT'S MOTHER CONTACTED DEXCOM ON (B)(6) 2015, TO REPORT THAT ON (B)(6) 2015 THE PATIENT'S RECEIVER HAD UNRECOVERABLE LOSS OF ANTENNA. THERE WAS NO REPORT OF INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. SUBSEQUENT TO THE INITIAL MDR, THE COMPLAINT RECEIVER WAS NOT RETURNED TO DEXCOM. THE TRANSMITTER (PART NUMBER STT-GL-003/SERIAL NUMBER (B)(4)/LOT NUMBER 5200720), BEING USED WITH THE COMPLAINT RECEIVER, WAS RETURNED ON (B)(4) 2015. THE RETURNED TRANSMITTER WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE TEST FAILED. THE REPORTED EVENT OF AN UNRECOVERABLE LOSS OF ANTENNA WAS CONFIRMED. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE TRANSMITTER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 760184 | DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM | MDS | MDS | DEXCOM, INC. | MT22608-PNK | 5201238 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 12 YR |