COMFORTDRIVE 200XDR
Report
- Report Number
- 3003637274-2015-00031
- Event Type
- Malfunction
- Date Received
- November 17, 2015
- Date of Event
- October 7, 2015
- Report Date
- November 2, 2015
- Manufacturer
- KAVO DENTAL GMBH
- Product Code
- EKX
- PMA / PMN Number
- K080677
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- DENTIST
Narratives
DURING A SHORT TEST RUN THE HEAT UP OF THE HANDPIECE HAS BEEN REPRODUCIBLE. ROOT CAUSE IS THAT THE BEARINGS HAVE BEEN GRITTY. THIS CAUSES HIGHER FRICTION AND THEREFORE INCREASE OF TEMPERATURE. ROOT CAUSE IS THAT THE BEARINGS ARE CONTENT OF A NORMAL WEAR PROCESS. THE HANDPIECE HAS NOT BEEN SENT IN FOR REPAIR SINCE THE PURCHASE IN (B)(6) 2011. THE ANALYSIS DID ALSO SHOW THAT THE PUSH BUTTON HAS CIRCULAR GROOVES ON THE INNER SIDE. THIS SHOWS THAT IT HAS BEEN PRESSED WHILE THE TOOL HAS BEEN SPINNING. ROOT CAUSE IS THAT THE HANDPIECE HAS BEEN USED AS A CHEEK RETRACTOR. THE IFU CONTAINS SEVERAL NOTES AND WARNINGS TO AVOID SUCH INCIDENTS: WARNING HAZARDS FOR THE CARE PROVIDER AND THE PATIENT. IN THE CASE OF DAMAGE, IRREGULAR RUNNING NOISE, EXCESSIVE VIBRATION, UNTYPICAL WARMING OR WHEN THE CUTTER OR GRINDER CANNOT BE HELD. DO NOT USE FURTHER AND NOTIFY SERVICE. CAUTION BURNING HAZARD FROM HOT INSTRUMENT HEAD OR HOT INSTRUMENTS COVER. IF THE INSTRUMENT OVERHEATS, BURNS MAY ARISE IN THE ORAL AREA. NEVER CONTACT SOFT TISSUE WITH THE INSTRUMENT HEAD OR INSTRUMENT COVER. CAUTION HAZARD FROM THE USE OF HANDPIECES EQUIPPED WITH ELECTRONIC MICROMOTORS. ELECTRONIC MICROMOTORS GENERATE MUCH MORE ENERGY THAN CONVENTIONAL PNEUMATIC TURBINES AND MOTORS. GIVEN THE HIGHER TORQUE AND SPEED, HANDPIECES THAT ARE POORLY SERVICED, DAMAGED OR USED IMPROPERLY CAN OVERHEAT WHICH CAN CAUSE SERIOUS BURN INJURIES TO THE PATIENT. OBSERVE THE FOLLOWING POINTS. THE FOLLOWING GUIDELINES MUST BE OBSERVED TO ENSURE SAVE USE OF THE ELECTRICALLY DRIVEN HANDPIECES: THE SERVICE INSTRUCTIONS FOR HANDPIECES MUST BE PRECISELY FOLLOWING WHEN USING KAVO SPRAY OR QUATTROCARE CARE SYSTEMS. BEFORE EACH USE, THE HANDPIECE MUST BE CHECKED FOR EXTERNAL DAMAGE. BEFORE EACH USE, PERFORM A TEST RUN WITH THE HANDPIECE, AND WATCH FOR ATYPICAL HEATING AND UNUSUAL NOISE AND VIBRATION. IMMEDIATELY STOP USING HANDPIECES THAT ACT UNUSUAL. NEVER PRESS THE PUSH BUTTON DURING OPERATION. THIS ALSO INCLUDES LIFTING THE CHEEK OR TONGUE! WE RECOMMEND RETURNING THE HANDPIECES TO KAVO AT REGULAR INTERVALS FOR TESTING, SETUP AND SERVICING. THE FREQUENCY OF THE CARE DEPENDS ON THE INSTRUMENTS USE. THE CONTRA-ANGLE HANDPIECES MUST BE SETUP ACCORDING TO THE KAVO INSTRUCTIONS TO ENSURE PROPER FUNCTIONING. (B)(4).
OVER A COUPLE OF MONTHS SEVERAL PATIENTS HAVE BEEN COMPLAINING THAT THE HANDPIECE WAS GETTING HOT TO LIP OR CHEEK DURING THE TREATMENT. NOBODY WAS INJURED. NOBODY IN THE DENTAL OFFICE DOES RECALL WHICH PATIENTS COMPLAINED ABOUT HEAT THEREFORE IT IS NOT POSSIBLE TO SUPPLY DETAILS ABOUT PATIENTS AND THE 'DATE OF EVENT' IS BEST ESTIMATE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 760302 | COMFORTDRIVE 200XDR | DENTAL HANDPIECE | EKX | KAVO DENTAL GMBH | 200XDR |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |