2.4MM TI MATRIXMANDIBLE LOCKING SCREW SLF-TPNG 10MM
Report
- Report Number
- 9612488-2015-10573
- Event Type
- Injury
- Date Received
- November 16, 2015
- Report Date
- October 30, 2015
- Manufacturer
- SYNTHES BETTLACH
- Product Code
- JEY
- PMA / PMN Number
- PK063790
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). (B)(6). DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
PRODUCT INVESTIGATION SUMMARY: THE FOLLOWING DEVICES WERE RECEIVED FOR EVALUATION: PART 04.503.638.01S / LOTS 9273378, 9369363, 9213139, 9273378, 9273378 (X5), PART 04.503.640.01S / LOT 9213139, 9257974 (X2), PART 04.503.642.01S / LOT 9219272 (X1), PART 04.503.644.01S / LOT 9279387 (X1). THE SHAFT AND HEAD THREADS AS WELL AS THE RECESS AND TIP SECTIONS SHOW SLIGHT MARKS OF USE. BASED ON THE SURGICAL TECHNIQUE, THE DEVICE SPECIFIC ADVERSE EVENTS OF LOOSENING, BENDING, OR BREAKAGE OF THE DEVICE CAN OCCUR. THE EXACT CAUSE FOR THE LOOSENING OF THE SCREWS COULD NOT BE DETERMINED. IT IS LIKELY THAT BONE MOVEMENT AND/OR INSUFFICIENT TIGHTENING OF THE SCREWS IN THE PLATE LED TO LOOSENING. NO MANUFACTURING-RELATED FAILURES WERE DETECTED. DEVICE USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. ADDITIONAL NARRATIVE: PATIENT INFORMATION NOT PROVIDED BY REPORTER. UNKNOWN WHEN DEVICE MALFUNCTIONED. NOT EXPLANTED. DEVICE IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION, BUT HAS YET TO BE RECEIVED. DEVICE HISTORY RECORDS WAS CONDUCTED. THE REPORT INDICATES THAT THE 04.503.640.01S / LOT 9213139: MANUFACTURING LOCATION: (B)(4), MANUFACTURING DATE: 22ND OCTOBER 2014, EXPIRY DATE: 1ST OCTOBER 2024, NO NCRS WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD (S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
THE DEVICE WAS RECEIVED, THE INVESTIGATION COULD NOT BE COMPLETED, AND NO CONCLUSION COULD BE DRAWN, AS PRODUCT IS ENTERING THE COMPLAINT SYSTEM. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
UPDATE: THE COMPLAINANT PART WAS RETURNED, INDICATING THAT A REVISION PROCEDURE WAS PERFORMED. NO ADDITIONAL INFORMATION WAS PROVIDED REGARDING THIS INFORMATION.
IT WAS FOUND ON (B)(6) 2015 THAT LOCKING SCREWS IMPLANTED IN THE PATIENT'S MANDIBLE HAVE BEEN LOOSENED. THE PATIENT HAD THE INITIAL SURGERY ON (B)(6) 2015. THIS REPORT IS 5 OF 9 FOR (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 755793 | 2.4MM TI MATRIXMANDIBLE LOCKING SCREW SLF-TPNG 10MM | BONE PLATE | JEY | SYNTHES BETTLACH | 9213139 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |