FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 5228043
·
Received November 16, 2015
Report
- Report Number
- 2032227-2015-66379
- Event Type
- Malfunction
- Date Received
- November 16, 2015
- Date of Event
- July 8, 2013
- Report Date
- November 4, 2015
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
FINDINGS: UNIT RECEIVED NO BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM. UNIT RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW. UNIT RECEIVED WITH CRACKED BATTERY TUBE THREADS. UNIT RECEIVED WITH CRACKED RESERVOIR TUBE LIP. UNIT RECEIVED WITH CRACKED CASE AT RESERVOIR TUBE WINDOW CORNER. UNIT RECEIVED MISSING ENDCAP STICKER. (B)(4).
Description of Event or Problem · 1
THE CUSTOMER REPORTED VIA PHONE CALL INDICATING THAT THE INSULIN PUMP ALARM BUTTON ERROR. CUSTOMER'S BLOOD GLUCOSE WAS UNKNOWN MG/DL. THE CUSTOMER IS UNABLE TO CLEAR THE ALARM. THE CUSTOMER PUMP IS OOE AND RETURNED TO BE SCRAPPED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 758132 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-522WWB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |