FDA Adverse Event Injury Summary report: N

BD WALMART RELION INSULIN SYRINGE

MDR report key: 5227342 · Received November 16, 2015

Report

Report Number
1920898-2015-00005
Event Type
Injury
Date Received
November 16, 2015
Date of Event
October 29, 2015
Report Date
November 4, 2015
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
K941657
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CATALOG #: OF NOTE, THE CATALOG NUMBER REPORTED BY THE CONSUMER WAS 311649. THIS CATALOG # IS FOR (B)(6) AND ITS COINCIDING BD CATALOG # IS 328507. SAMPLES, 2 USED LOOSE AND 80 UNUSED IN SEALED POLYBAGS, WERE RETURNED FOR EVALUATION. A VISUAL ANALYSIS OF THE USED UNITS REVEALED BROKEN CANNULAS. A MICROSCOPIC ANALYSIS OF THE USED UNITS REVEALED CHARACTERISTICS SUCH AS RESIDUAL BENDS ON THE BROKEN HUB END, CRACKED ADHESIVE, AND TUBING OVALITY (DEFORMATION FROM THE NORMALLY CIRCULAR CROSS SECTION). REMOVAL OF SOME OF THE EPOXY MADE THIS OBSERVATION MORE APPARENT. THIRTY OUT OF 80 SAMPLES IN THE SEALED POLY BAGS WERE ALSO EXAMINED AND NO BENT OR BROKEN CANNULA WAS OBSERVED. A REVIEW OF THE DEVICE HISTORY RECORD REVEALED NO IRREGULARITIES DURING THE MANUFACTURE OF THE REPORTED LOT NUMBER 5033899. CONCLUSION: AN ABSOLUTE ROOT CAUSE FOR THIS INCIDENT CANNOT BE DETERMINED. HOWEVER, OUR QUALITY ENGINEER NOTES THAT WHEN VIEWED TOGETHER, ALL OF THE INVESTIGATION FINDINGS ARE INDICATORS OF BENDING/RE-STRAIGHTENING MODE OF FAILURE. DEVICE NOT RETURNED FOR EVALUATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT A CONSUMER HAD TWO BD (B)(6) RELION INSULIN SYRINGE NEEDLES BREAK OFF IN HER ABDOMEN WHILE SHE WAS INJECTING HERSELF. THE CONSUMER WENT TO AN EMERGENCY DEPARTMENT TO BE EVALUATED AND RECEIVED AN X-RAY. THE X-RAY DID NOT VISUALIZE THE NEEDLES AND SHE REPORTED THAT SHE WAS ADVISED TO WAIT AND SEE IF THE INJECTION SITE BECOMES INFECTED. NO FURTHER MEDICAL INTERVENTIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
755756 BD WALMART RELION INSULIN SYRINGE INSULIN SYRINGE FMF BD MEDICAL - DIABETES CARE 5033899

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention