FDA Adverse Event
Other
Summary report: N
AMBULIFT C-3
MDR report key: 52210
·
Received November 25, 1996
Report
- Report Number
- 52210
- Event Type
- Other
- Date Received
- November 25, 1996
- Date of Event
- November 13, 1996
- Report Date
- November 25, 1996
- Manufacturer
- ARJO-CENTURY INC.
- Product Code
- FSA
- Adverse Event
- Yes
- Report Source
- User Facility report
- Reporter Location
- NE, US
- Reporter Occupation
- NURSE
Narratives
Description of Event or Problem · 1
THIS RESIDENT BEING TRANSFERRED FROM WHEELCHAIR TO BED WITH C-3 LIFT. BEGAN TO TIP/SLIDE BACKWARDS, HEAD FIRST. NURSING ASST CAUGHT THE RESIDENT BUT HE DID HIT HEAD ON FLOOR. MINIMAL BLEEDING FROM NICKEL-SIZE AREA ON BACK OF HEAD. AFTER ANALYSIS OF ACCIDENT THE STAFF DISCOVERED THAT POSITIONING OF THE SLING BAR DETERMINES BALANCE OF THE RESIDENT, AND THAT THE HOOK OF THE SLING BAR MUST BE AWAY FROM THE RESIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AMBULIFT C-3 | MECHANICAL SLING LIFT | FSA | ARJO-CENTURY INC. | C-3 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 87 YR | Other |