FDA Adverse Event Malfunction Summary report: N

SILVERHAWK LX-M

MDR report key: 5219381 · Received November 12, 2015

Report

Report Number
2183870-2015-00449
Event Type
Malfunction
Date Received
November 12, 2015
Date of Event
April 29, 2014
Report Date
April 30, 2014
Manufacturer
COVIDIEN
Product Code
MCW
PMA / PMN Number
K053460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS MDR IS BEING SUBMITTED AS A PART OF A RETROSPECTIVE REVIEW / REMEDIATION EFFORT PERFORMED AT THE COVIDIEN (B)(4) LOCATION, FOLLOWING MEDTRONIC¿S ACQUISITION OF COVIDIEN. A CAPA HAS BEEN OPENED TO MANAGE THE ACTIONS RELATED TO REMEDIATION OF COMPLAINT FILES AND ANY REQUIRED MDR REPORTING. (B)(4). (B)(6).

Description of Event or Problem · 1

IT IS REPORTED THAT DURING THE PROCEDURE THE PHYSICIAN MADE MULTIPLE PASSES. DIFFICULTY WAS NOTED WHEN TRYING TO ADVANCE IN A CERTAIN PLANE. THE PHYSICIAN THEN EXPERIENCED DIFFICULTY TURNING THE CUTTER OFF SO THE DEVICE WAS REMOVED. DIFFICULTY WAS EXPERIENCED WHEN TRYING TO CLEAN THE DEVICE. THE TISSUE PLUNGER WAS FORCED FORWARD BUT THE PHYSICIAN FELT THIS MAY HAVE DAMAGED THE DEVICE SO A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT INJURY IS REPORTED. EVALUATION SUMMARY: THE SILVERHAWK WAS RECEIVED FOR EVALUATION. THE DISTAL TIP ASSEMBLY EXHIBITED A RUPTURE OF THE HOUSING PET, COIL, TECOTHANE WALL. THE RUPTURE CONTAINED CALCIFIC TISSUE RESIDUE THAT HAD ERUPTED THROUGH THE TIP ASSEMBLY COMPONENTS. THE PLUNGER COMPONENT WAS LOCATED IMMEDIATELY DISTAL TO THE COLLECTED TISSUE. SOME OF THE ERUPTED TISSUE WAS PRIED OFF THE ASSEMBLY REVEALING LASER-CUT COIL STRANDS AND POLYMER FILM. THE HOUSING'S PLATINUM-IRIDIUM LASER-CUT COIL SECTION EXHIBITED SEVERAL DEFORMATIONS INCLUDING SOME CUTTING AT THE DISTAL EDGE OF THE HOUSING WINDOW AND CUTTING AND DELAMINATION OF THE HOUSING'S LASER-CUT COIL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
750298 SILVERHAWK LX-M CATHETER, PERIPHERAL, ATHERECTOMY MCW COVIDIEN P4055 9828046

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention