OT VERIO2 METER
Report
- Report Number
- 2939301-2015-48137
- Event Type
- Malfunction
- Date Received
- November 10, 2015
- Report Date
- October 30, 2015
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K131363
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AU
- Reporter Occupation
- PATIENT
Narratives
FOLLOW-UP # 1: THE LAY USER/PATIENTS TEST STRIPS HAVE BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE TEST STRIPS INVOLVED WITH THIS COMPLAINT FAILED TESTING. THE REPORTED ISSUE WAS CONFIRMED. ADDITIONAL TESTS WERE PERFORMED ON THE TEST STRIP VIAL AND THE DESICCANT TO CHECK THE STRUCTURAL INTEGRITY AND FOR POSSIBLE MOISTURE ISSUE. THE STRUCTURAL INTEGRITY OF THE TEST STRIP VIAL WAS FOUND TO BE INTACT DURING A GROSS LEAK TEST. THE DESICCANT WITHIN THE SUBJECT VIAL WAS FOUND TO CONTAIN MORE MOISTURE THAN EXPECTED FROM NORMAL USE (AS PER PRODUCT LABELLING). IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.
ON (B)(6) 2015, THE REPORTER CONTACTED LIFESCAN (B)(4), ALLEGING INACCURATE HIGH RESULTS OF "316, 358 AND 330 MG/DL" ON THE SUBJECT METER COMPARED TO "182, 308 AND 171 MG/DL" ON A ONETOUCH ULTRA METER, PERFORMED WITHIN 30 MINUTES OF EACH OTHER. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED RESULTS DID NOT MEET LIFESCAN'S ACCURACY CRITERIA AND THE ALLEGED INACCURACY ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 743253 | OT VERIO2 METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR |