DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM
Report
- Report Number
- 3004753838-2015-66433
- Event Type
- Malfunction
- Date Received
- November 9, 2015
- Date of Event
- October 15, 2015
- Report Date
- October 15, 2015
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
(B)(4).
(B)(4).
PATIENT'S MOTHER CONTACTED DEXCOM ON (B)(6) 2015 TO REPORT NO AUDIO OUTPUT ON (B)(6) 2015. AT THE TIME OF CONTACT THE PATIENT'S MOTHER DID NOT REPORT ANY INJURY OR MEDICAL INTERVENTION. THE DEVICE HAS BEEN RECEIVED FOR EVALUATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE THE EVALUATION IS COMPLETE.
THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS EXTERNALLY VISUALLY INSPECTED AND NO DEFECT WAS FOUND. A REVIEW OF THE DOWNLOADED RECEIVER LOG DID NOT FIND ANY ISSUES RELATED TO THE CUSTOMER COMPLAINT. FUNCTIONAL TESTING AND A MANUAL DROP TEST FOR INTERMITTENCY WAS PERFORMED AND THE TESTS FAILED. THE RECEIVER CASE WAS OPENED FOR FURTHER EVALUATION. A VISUAL INTERIOR INSPECTION WAS PERFORMED AND THE INSPECTION PASSED. A SPEAKER RESISTANCE TEST WAS PERFORMED AND CONFIRMED THE REPORTED EVENT OF NO AUDIO OUTPUT. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE SPEAKER ASSEMBLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 742107 | DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM | MDS | MDS | DEXCOM, INC. | MT22608-PNK | 5202415 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 6 YR |