DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2015-62992
- Event Type
- Malfunction
- Date Received
- November 9, 2015
- Date of Event
- October 18, 2015
- Report Date
- October 18, 2015
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- UDI-DI
- 00386270000125
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
(B)(4).
(B)(4).
THE COMPLAINT TRANSMITTER WAS NOT RETURNED TO DEXCOM. THE RECEIVER (PART NUMBER STK-PR-PNK/SERIAL NUMBER (B)(4)/LOT NUMBER 5201273), BEING USED WITH THE COMPLAINT TRANSMITTER, WAS RETURNED ON (B)(4) 2015. THE RETURNED RECEIVER WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE REPORTED FAULT COULD NOT BE REPRODUCED AND THERE WAS NO FAILURE DETECTED. THE DEVICE WAS DETERMINED TO BE OPERATING WITHIN THE REQUIRED SPECIFICATIONS WITHOUT MALFUNCTION; HOWEVER, A REVIEW OF THE DIAGNOSTIC CHART CONFIRMED THE REPORTED EVENT OF AN INTERMITTENT OUT OF RANGE SIGNAL. A ROOT CAUSE COULD NOT BE DETERMINED.
PATIENT'S MOTHER CONTACTED DEXCOM ON (B)(6) 2015 TO REPORT THAT ON (B)(6) 2015, PATIENT EXPERIENCED AN INTERMITTENT OUT OF RANGE SIGNAL. THE COMPLAINT DEVICE WAS NOT RETURNED FOR EVALUATION. HOWEVER, THE DATA LOG WAS PROVIDED BY THE PATIENT. THE DATA WAS REVIEWED AND DID NOT CONFIRM THE REPORTED EVENT OF INTERMITTENT ANTENNA ICON. A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 742045 | DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS | MDS | DEXCOM, INC. | 9438-05 | 5203773 | 00386270000125 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 7 YR |