VASOVIEW 7 XB
Report
- Report Number
- 2242352-2015-01244
- Event Type
- Malfunction
- Date Received
- November 6, 2015
- Date of Event
- October 14, 2015
- Report Date
- October 14, 2015
- Manufacturer
- MAQUET CV
- Product Code
- GEI
- PMA / PMN Number
- K041981
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). THE DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION. EVIDENCE OF BLOOD AND SIGNS OF SOME CLINICAL USE WERE OBSERVED ON THE CANNULA. THE SCOPE WASH TUBING WAS KINKED AND WAS TRANSECTED IMMEDIATELY BELOW THE ADHESIVE JOINT TO THE C-RING. ADHESIVE WAS OBSERVED AT THE JOINT. THE SCOPE WASH TUBING REMAINED ATTACHED TO THE CANNULA DUE THE PRESENCE OF A RETENTION RIB. THE BOND BETWEEN THE WIRE AND THE C-RING REMAINED INTACT. DURING VISUAL INSPECTION, COMPRESSED AREAS OF DEFORMATION AND BENDING WERE OBSERVED. THE TRANSECTED SURFACES EXHIBITED TEARING AND PLASTIC DEFORMATION CONSISTENT WITH BENDING AND PULLING WHICH RESULTED IN THE TUBING BREAK. THE TUBING WAS PULLED FORWARD IN THE DIRECTION OF TOOL TRAVEL AND DISPLAYED KINKS LOCATED IN THE TOOL PATH. BASED ON THE CONDITION OF THE DEVICE AS RETURNED, WE WERE ABLE TO CONFIRM THE REPORTED COMPLAINT FOR ¿C-RING ASSEMBLY/SCOPE WASH TUBING TRANSECTED BY CAUTERY TOOL¿. THE ROOT CAUSE IS ENGAGEMENT OF THE SCOPE WASH TUBING WITH THE FORCEPS ON THE CAUTERY TOOL DURING USE. THE ROOT CAUSE IS ADDRESSED IN THE INSTRUCTIONS FOR USE. THE IFU PROVIDES INSTRUCTION TO THE USER; "TAKE CARE TO AVOID ENGAGING THE COMPONENTS OF THE C-RING ASSEMBLY WITH THE FORCEPS. DOING SO MAY DAMAGE PORTIONS OF THE C-RING ASSEMBLY CAUSING THE COMPONENTS TO SEPARATE FROM THE ASSEMBLY AND FALL INTO THE BODY." THIS DEVICE LOT HAS BEEN MANUFACTURED WITH THE ADDITION OF THE RETENTION RIB TO PREVENT THE TRANSECTED PORTION OF THE SCOPE WASH TUBING ON THE C-RING ASSEMBLY FROM DISLODGING INTO THE PATIENT. (B)(4). THE DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION. EVIDENCE OF BLOOD AND SIGNS OF SOME CLINICAL USE WERE OBSERVED ON THE CANNULA. THE SCOPE WASH TUBING WAS KINKED AND WAS TRANSECTED IMMEDIATELY BELOW THE ADHESIVE JOINT TO THE C-RING. ADHESIVE WAS OBSERVED AT THE JOINT. THE SCOPE WASH TUBING REMAINED ATTACHED TO THE CANNULA DUE THE PRESENCE OF A RETENTION RIB. THE BOND BETWEEN THE WIRE AND THE C-RING REMAINED INTACT. DURING VISUAL INSPECTION, COMPRESSED AREAS OF DEFORMATION AND BENDING WERE OBSERVED. THE TRANSECTED SURFACES EXHIBITED TEARING AND PLASTIC DEFORMATION CONSISTENT WITH BENDING AND PULLING WHICH RESULTED IN THE TUBING BREAK. THE TUBING WAS PULLED FORWARD IN THE DIRECTION OF TOOL TRAVEL AND DISPLAYED KINKS LOCATED IN THE TOOL PATH. BASED ON THE CONDITION OF THE DEVICE AS RETURNED, WE WERE ABLE TO CONFIRM THE REPORTED COMPLAINT FOR ¿C-RING ASSEMBLY/SCOPE WASH TUBING TRANSECTED BY CAUTERY TOOL¿. THE ROOT CAUSE IS ENGAGEMENT OF THE C-RING WITH THE FORCEPS ON THE CAUTERY TOOL DURING USE. THE ROOT CAUSE IS ADDRESSED IN THE INSTRUCTIONS FOR USE. THE IFU PROVIDES INSTRUCTION TO THE USER; "TAKE CARE TO AVOID ENGAGING THE COMPONENTS OF THE C-RING ASSEMBLY WITH THE FORCEPS. DOING SO MAY DAMAGE PORTIONS OF THE C-RING ASSEMBLY CAUSING THE COMPONENTS TO SEPARATE FROM THE ASSEMBLY AND FALL INTO THE BODY." THIS DEVICE LOT HAS BEEN MANUFACTURED WITH THE ADDITION OF THE RETENTION RIB TO PREVENT THE TRANSECTED PORTION OF THE SCOPE WASH TUBING ON THE C-RING ASSEMBLY FROM DISLODGING INTO THE PATIENT.
NOVEMBER 05, 2015 06:12 PM (GMT-5:00) ADDED BY (B)(6): A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NONCONFORMANCE RECORDED IN THE LOT HISTORY. THE DEVICE HAS NOT YET BEEN RETURNED TO MAQUET CARDIAC SURGERY FOR EVALUATION. WE ARE FOLLOWING UP WITH THE CUSTOMER FOR THE RETURN OF THE DEVICE. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF THE DEVICE IS RECEIVED.
THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE VASOVIEW 7 C-RING BECAME DETACHED DURING USE. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS.
THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE VASOVIEW 7 C-RING BECAME DETACHED DURING USE. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 737831 | VASOVIEW 7 XB | ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES | GEI | MAQUET CV | 25108961 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |