1823260-2015-04450
Report
- Report Number
- 1823260-2015-04450
- Event Type
- Malfunction
- Date Received
- November 5, 2015
- Date of Event
- October 16, 2015
- Report Date
- December 1, 2015
- PMA / PMN Number
- K961481
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TH
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
A SPECIFIC ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE AVAILABLE INFORMATION, A SAMPLE HANDLING ISSUE CANNOT BE EXCLUDED AS THE CENTRIFUGATION TIME WAS VERY SHORT (3 MINUTES).
THIS EVENT OCCURRED IN (B)(6). (B)(4)
BASED ON A REVIEW OF THE ALARM TRACE, THERE IS NO INDICATION OF AN INSTRUMENT ISSUE. THE ANALYZER PERFORMANCE CHECK ON (B)(6) 2015 WAS ACCEPTABLE.
THE CUSTOMER COMPLAINED OF ERRONEOUS RESULTS FOR 1 PATIENT TESTED FOR TROPONIN T HS STAT (HIGH SENSITIVE SHORT TURN AROUND TIME) TROPONIN T HS STAT. THE ERRONEOUS RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE INITIAL TROPONIN T HS STAT RESULT WAS 169.2 PG/ML. THIS RESULT WAS REPORTED OUTSIDE OF THE LABORATORY. THIS RESULT WAS BELIEVED TO BE TOO HIGH. A SECOND SAMPLE WAS OBTAINED AND THE RESULT WAS 3.00 PG/ML WITH A DATA FLAG. A THIRD SAMPLE WAS OBTAINED AND THE RESULT WAS 3.00 PG/ML WITH A DATA FLAG. THE INITIAL SAMPLE WAS REPEATED AND THE RESULT WAS 3.00 PG/ML WITH A DATA FLAG. NO ADVERSE EVENT WAS REPORTED. THE TROPONIN T HS STAT REAGENT LOT WAS 182603. THE EXPIRATION DATE WAS NOT PROVIDED. CALIBRATION AND QUALITY CONTROLS WERE ACCEPTABLE. IT WAS NOTED THAT THE CENTRIFUGATION TIME WAS 3 MINUTES. THIS IS A VERY SHORT TIME AS THE RECOMMENDED CENTRIFUGATION TIME DEPENDING ON TUBE MANUFACTURERS IS BETWEEN 10-15 MINUTES. A SPECIFIC ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE DATA PROVIDED, A SAMPLE HANDLING ISSUE CANNOT BE EXCLUDED.
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |