PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2015-63104
- Event Type
- Malfunction
- Date Received
- November 4, 2015
- Date of Event
- October 18, 2015
- Report Date
- October 18, 2015
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA
- Reporter Occupation
- OTHER
Narratives
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).
THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE AND ALARMED FOR A BUTTON ERROR DUE TO CORRODED KEYPAD TRACES. NO MOISTURE DAMAGE WAS NOTED TO THE LIQUID CRYSTAL DISPLAY CIRCUIT BOARD, MOTHER BOARD, INTERFACE BOARD, RADIO FREQUENCY BOARD, MOTOR, VIBRATOR OR BATTERY TUBE ASSEMBLY DURING THE VISUAL INSPECTION. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED CASE AT THE DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHES ON THE DISPLAY WINDOW AND CRACKED BATTERY TUBE THREADS.
THE CUSTOMER REPORTED VIA PHONE CALL THAT HER INSULIN PUMP MADE A LOUD BEEPING SOUND AND THEN NOTICED THAT THERE WAS MOISTURE ON THE PUMP. TROUBLESHOOTING WAS INITIATED FOR THE PUMP EXPOSURE TO FLUID. THE PATIENT'S BLOOD GLUCOSE AT THE TIME OF INCIDENT IS UNKNOWN. THE CUSTOMER WAS WEARING HER INSULIN PUMP IN HER BRA WHILE AT CHURCH; SHE BELIEVES THAT SHE SWEATED ON IT. THE CONSTANT TONE WAS UNABLE TO BE CLEARED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS AND A REPLACEMENT PUMP WAS SHIPPED TO THE CUSTOMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 733546 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-523NAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR |