FDA Adverse Event Injury Summary report: N

SYNERGY¿

MDR report key: 5202297 · Received November 4, 2015

Report

Report Number
2134265-2015-07497
Event Type
Injury
Date Received
November 4, 2015
Date of Event
October 9, 2015
Report Date
October 13, 2015
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
NIQ
PMA / PMN Number
SIMILAR
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATED BY MFR: THE COMPLAINT DEVICE WAS NOT RETURNED FOR ANALYSIS. THE BATCH NUMBER IS UNKNOWN AND THE MANUFACTURING RECORDS FOR THE COMPLAINT DEVICE COULD NOT BE REVIEWED. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Additional Manufacturer Narrative · 1

BRAND NAME CORRECTED FROM SYNERGY BALLOON DILATATION CATHETER TO SYNERGY¿. COMMON DEVICE NAME CORRECTED FROM CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL TO CORONARY DRUG-ELUTING STENT. UPN- SEARCH CORRECTED FROM UNK425 - SYNERGY - CMC - (B)(4) (PERIPHERAL INTERV) - 35000 TO UNK776 - SYNERGY II DES - CMC - (B)(4) (INTERVENT CARDIO) - 30000. UPN CORRECTED FROM UNK425 TO UNK776. PMA# OR 510K# CORRECTED FROM K993305 TO SIMILAR. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT STENT DISLODGEMENT INSIDE THE PATIENT OCCURRED. A SYNERGY DRUG ELUTING STENT WAS ADVANCED TO TREAT THE LESION. HOWEVER, DURING PROCEDURE, THE STENT CAME OFF THE SHAFT AND EMBOLIZED. THE PROCEDURE WAS COMPLETED WITH A NON BSC STENT. NO PATIENT COMPLICATIONS WERE REPORTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT STENT DISLODGEMENT INSIDE THE PATIENT OCCURRED. A SYNERGY DRUG ELUTING STENT WAS ADVANCED TO TREAT THE LESION. HOWEVER, DURING PROCEDURE, THE STENT CAME OFF THE SHAFT AND EMBOLIZED. THE PROCEDURE WAS COMPLETED WITH A NON BSC STENT. NO PATIENT COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
732050 SYNERGY¿ CORONARY DRUG-ELUTING STENT NIQ BOSTON SCIENTIFIC - GALWAY UNK776

Patients

Seq Age Sex Outcome Treatment
1 Other