TRILOGY 100

MDR report key: 5200955 · Received November 4, 2015

Report

Report Number
2518422-2015-03592
Event Type
Malfunction
Date Received
November 4, 2015
Date of Event
October 9, 2015
Report Date
October 9, 2015
Manufacturer
RESPIRONICS, INC.
Product Code
CBK
UDI-DI
00606959015364
PMA / PMN Number
K083526
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Description of Event or Problem · 1

THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR WAS ALARMING FOR A "HIGH TEMPERATURE" ALARM CONDITION. THERE WAS NO HARM OR INJURY REPORTED. THE VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVICE'S BLOWER MOTOR AND OUTLET FLOW PATH THERMISTOR WERE REPLACED TO ADDRESS THE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
731352 TRILOGY 100 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS, INC. 1054260 00606959015364