FDA Adverse Event Malfunction Summary report: N

STABLE BASE

MDR report key: 5198164 · Received October 29, 2015

Report

Report Number
8020045-2015-00064
Event Type
Malfunction
Date Received
October 29, 2015
Report Date
October 29, 2015
Manufacturer
LEONHARD LANG GMBH
Product Code
DRX
PMA / PMN Number
K024247
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE RETAINED SAMPLES OF THE COMPLAINED LOT HAVE BEEN INSPECTED VISUALLY AND TESTED ELECTRICALLY. THEY FAILED THE ELECTRICAL TESTS. THE ROOT CAUSE OF THIS DEFECT IS STILL UNDER INVESTIGATION. HOWEVER, IN THE COURSE OF THE INVESTIGATION CONCERNING MDR 8020045-2015-0048, IT WAS RECENTLY DETERMINED THAT SEVERAL MODELS SHARING A CERTAIN DESIGN AND MANUFACTURING PROCESS HAVE A HIGH PROBABILITY TO DEVELOP A DEFECT. WE HAVE THEREFORE DECIDED TO PERFORM A RECALL OF THESE MODELS. BASED ON THIS DETERMINATION WE HAVE REEVALUATED THIS COMPLAINT AND DO NOW BELIEVE IT IS LINKED ALTHOUGH IT WAS NOT INITIALLY OBVIOUS. WE HAVE THEREFORE DECIDED TO REPORT THIS INCIDENT. ANY FURTHER FINDINGS IN THE INVESTIGATION WILL BE REPORTED AS FOLLOW-UP REPORTS TO MDR 8020045-2015-00048 ONLY.

Description of Event or Problem · 1

ON (B)(6) 2015, WE HAVE BEEN INFORMED ABOUT AN INCIDENT INVOLVING ECG ELECTRODES. THE DISTRIBUTOR COMPLAINED THAT A "CUSTOMER CALLED ABOUT THE SW620 ELECTRODES BEING DRIED OUT. THE LOT NUMBER IS 40730-0324. NOT ALL BAGS ARE DRIED OUT AND THE CUSTOMER SAYS "IT IS HIT OR MISS".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
720827 STABLE BASE DRX LEONHARD LANG GMBH W-60 40730-0324

Patients

Seq Age Sex Outcome Treatment
1 UNK