Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION:THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/19/2015 WITH THE FOLLOWING FINDINGS:THE KEYPAD WAS FOUND TO BE PEELING AND PUNCTURED. EVALUATION REVEALED ALL KEYPAD BUTTONS WERE APPROPRIATELY RESPONSIVE. THERE WAS EVIDENCE OF KEYPAD CONTAMINATION FOUND UNDER ALL KEY CONTACTS. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED KEYPAD BUTTON CONTACT CONTAMINATION UNDER ALL CONTACTS. THIS REPORT IS MADE BASED ON RESULTS OF THE INVESTIGATION PERFORMED ON 10/19/2015.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
725308 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 59 YR