FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 5190259 · Received October 30, 2015

Report

Report Number
2032227-2015-61445
Event Type
Malfunction
Date Received
October 30, 2015
Date of Event
October 13, 2015
Report Date
October 13, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED MOTOR ERROR DURING THE OCCLUSION TEST AND UNABLE TO PRIME DURING PRIME TEST DUE TO FAULTY FORCE SENSOR RESISTOR HOWEVER, THE MOTOR PASSED MOTOR TEST. THE INSULIN PUMP WAS RECEIVED WITH CRACKED CASE AT THE DISPLAY WINDOW CORNER AND MINOR SCRATCHED DISPLAY WINDOW.

Description of Event or Problem · 1

THE CUSTOMER'S WIFE CALLED IN TO REPORT A MOTOR ERROR ALARM. THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS UNKNOWN. THE CALLER COMPLETED TROUBLESHOOTING AND WAS ABLE TO REWIND AND FILL THE TUBING. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. THE CUSTOMER WAS INFORMED THAT THE DEVICE WOULD BE REPLACED, AND WAS INFORMED TO RETURN THE MALFUNCTIONING DEVICE BACK FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
722434 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 51 YR