FDA Adverse Event Malfunction Summary report: N

DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM

MDR report key: 5188401 · Received October 29, 2015

Report

Report Number
3004753838-2015-62758
Event Type
Malfunction
Date Received
October 29, 2015
Date of Event
October 2, 2015
Report Date
October 6, 2015
Manufacturer
DEXCOM, INC.
Product Code
MDS
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

THE COMPLAINT RECEIVER WAS NOT RETURNED TO DEXCOM. THE TRANSMITTER (PART NUMBER STT-GL-003/SERIAL NUMBER (B)(4)/LOT NUMBER 5200716), BEING USED WITH THE COMPLAINT RECEIVER, WAS RETURNED ON 11/20/2015. THE RETURNED TRANSMITTER WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE TEST FAILED. THE REPORTED EVENT OF AN UNRECOVERABLE LOSS OF ANTENNA HAS BEEN CONFIRMED. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE TRANSMITTER.

Description of Event or Problem · 1

PATIENT'S MOTHER CONTACTED DEXCOM ON BEHALF OF THE PATIENT ON (B)(6) 2015, TO REPORT THAT ON (B)(6) 2015 THE PATIENT'S RECEIVER WAS OUT OF RANGE FOR 24 HOURS. THERE WAS NO REPORT OF INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
718923 DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM MDS MDS DEXCOM, INC. MT22608-BLU 5203292

Patients

Seq Age Sex Outcome Treatment
1 6 YR