ACTIVA
Report
- Report Number
- 3007566237-2015-03138
- Event Type
- Malfunction
- Date Received
- October 29, 2015
- Report Date
- October 6, 2015
- Manufacturer
- MEDTRONIC NEUROMODULATION
- Product Code
- MHY
- PMA / PMN Number
- P960009
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
CONCOMITANT MEDICAL PRODUCTS: PRODUCT ID 7426, SERIAL# (B)(4), IMPLANTED: (B)(6) 2007, EXPLANTED: (B)(6) 2015, PRODUCT TYPE IMPLANTABLE NEUROSTIMULATOR. PRODUCT ID 7482, SERIAL# (B)(4), IMPLANTED: (B)(6) 2007, EXPLANTED: (B)(6) 2015, PRODUCT TYPE EXTENSION. (B)(4).
A HEALTH CARE PROVIDER (HCP) REPORTED VIA A MANUFACTURING REPRESENTATIVE THAT IN 2014 THE PATIENT HAD LESS DEEP BRAIN STIMULATION (DBS) EFFECT. RECENTLY A FRACTURE IN THE SYSTEM WAS SUSPECTED DUE TO UNSTABLE IMPEDANCES. ABNORMAL IMPEDANCES WERE MEASURED AND PRIOR TO THE REVISION, IMPEDANCES OF ELECTRODE THREE ON THE LEFT SIDE WAS UNSTABLE. THE IMPEDANCES VALUES IMPLIED A SHORT OR A FRACTURE. THERE WERE NO SUSPICIOUS FINDINGS ON X-RAYS. A REVISION TO REPLACE THE EXTENSION AND IMPLANTABLE NEUROSTIMULATOR (INS) WAS DONE. DURING THE REVISION, BLOOD WAS FOUND TO BE MIXED IN THE UPPER PORTION OF THE INS WERE ELECTRODES 0-3 WERE LOCATED AND ONLY A HALF OF THE BOOT WAS COVERING THE LEAD AND ADAPTOR. AFTER THE REVISION, IMPEDANCES OF ELECTRODE ZERO WERE MEASURED TO BE HIGH. THE CAUSE OF THE HIGH IMPEDANCES, NO INTERVENTIONS, AND NO OUTCOME WAS REPORTED REGARDING THIS EVENT. FURTHER FOLLOW-UP IS BEING CONDUCTED TO OBTAIN THIS INFORMATION. IF ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP REPORT WILL BE SENT. REFER TO MANUFACTURER REPORT #9614453-2015-02686.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 719636 | ACTIVA | STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR | MHY | MEDTRONIC NEUROMODULATION | 37602 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |