FDA Adverse Event Malfunction Summary report: N

VMAX229 SERIES PULMONARY/METABOLIC SYSTEMS

MDR report key: 5186011 · Received October 28, 2015

Report

Report Number
2021710-2015-02075
Event Type
Malfunction
Date Received
October 28, 2015
Date of Event
September 17, 2015
Report Date
October 12, 2015
Manufacturer
CAREFUSION
Product Code
BTY
PMA / PMN Number
K942211
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
RESPIRATORY THERAPIST

Narratives

Additional Manufacturer Narrative · 1

AN EVALUATION WAS PERFORMED BY A CAREFUSION FAILURE ANALYSIS TECHNICIAN ON THE DELL XE2 COMPUTER. THE REPORTED COMPLAINT WAS DUPLICATED BY POWERING UP THE COMPUTER AND THE SCREEN SHOWED ¿WINDOWS COULD NOT START BECAUSE OF A COMPUTER DISK HARDWARE CONFIGURATION PROBLEM.¿ THIS BEHAVIOR WAS FOUND TO OCCUR IF THE DEFAULT VALUE IN THE BOOT.INI FILE IS MISSING OR INVALID, WINDOWS XP IS NOT INSTALLED IN THE LOCATION SPECIFIED IN THE BOOT.INI FILE, THE NTOSKRNL.EXE FILE IS MISSING OR DAMAGED, THE PARTITION PATH IN THE BOOT.INI FILE IS NOT SET CORRECTLY OR BECAUSE OF A GENERAL HARDWARE FAILURE. IN CONCLUSION, IT WAS DETERMINED THAT THERE WAS AN ISSUE WITH THE HARDWARE CONFIGURATION.

Additional Manufacturer Narrative · 1

(B)(4). ONCE THE DEVICE IS RETURNED FOR EVALUATION AND ADDITIONAL INFORMATION IS RECEIVED A FOLLOW UP WILL BE SUBMITTED. CAREFUSION CONTINUES TO TRACK AND TREND ANY INCIDENT RELATED TO THIS ISSUE. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED EXPERIENCING COMPUTER FREEZE-UPS THAT WERE OCCURRING 2-3 TIMES A WEEK. DURING A TREADMILL TEST, ONE OF THEIR PROTOCOL SETTINGS WAS NOT TRANSFERRED CORRECTLY. THE TREADMILL SPEED WAS SUPPOSED TO BE SET AT 0.6 MPH AND IT JUMPED UP TO 2.5 MPH ON AN ELDERLY LADY. SHE DID NOT FALL NOR WAS THERE ANY INCIDENT RESULTING FROM THIS AS THE THERAPIST WAS ABLE TO MANUALLY DECREASE THE SPEED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
713094 VMAX229 SERIES PULMONARY/METABOLIC SYSTEMS CALCULATOR, PREDICTED VALUES, PULMONARY FUNCTION BTY CAREFUSION 229 E

Patients

Seq Age Sex Outcome Treatment
1 86 YR