FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 5183516 · Received October 28, 2015

Report

Report Number
2531779-2015-38931
Event Type
Malfunction
Date Received
October 28, 2015
Report Date
October 15, 2015
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6)2015 WITH THE FOLLOWING FINDINGS: DURING TESTING, THE OK KEYPAD BUTTON WAS INTERMITTENTLY UNRESPONSIVE. THE KEYPAD COVER WAS REMOVED AND THE OK KEY CONTACT WAS FOUND TO BE INVERTED. EVIDENCE OF CONTAMINATION WAS FOUND UNDER THE OK BUTTON CONTACT. THE BATTERY COMPARTMENT WAS FOUND TO BE CRACKED ON THE LEFT SIDE. ADDITIONALLY, THE KEYPAD COVER WAS MISSING FROM THE PUMP. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A KEYPAD BUTTON RESPONSE ISSUE WITH A BUTTON CONTACT DEFECT AND EVIDENCE OF CONTAMINATION. ADDITIONALLY, THE BATTERY COMPARTMENT WAS FOUND TO BE DAMAGED. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(6) 2015.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
712919 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 29 YR