FDA Adverse Event Malfunction Summary report: N

SELF-DRILLING HALF PIN APEX Ø 3MM, 110 X 25MM

MDR report key: 5183296 · Received October 28, 2015

Report

Report Number
0008031020-2015-00440
Event Type
Malfunction
Date Received
October 28, 2015
Date of Event
October 9, 2015
Report Date
October 9, 2015
Manufacturer
STRYKER GMBH
Product Code
JDW
PMA / PMN Number
K861766
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED INCIDENT THAT SELF-DRILLING HALF PIN APEX Ø 3MM, 110 X 25MM WAS ALLEGEDLY BROKEN DURING INSERTION COULD BE CONFIRMED. BASED ON INVESTIGATION, THE ROOT CAUSE WAS ATTRIBUTED TO A R&D/DESIGN RELATED ISSUE. THE FAILURE WAS CAUSED BY TOO HIGH TOLERANCES ON THE TIP OF THE PIN. AN NC AND CAPA HAVE BEEN OPEN TO ADDRESS THE RECURRING SITUATION OF THE SELF-DRILLING HALF PIN APEX Ø 3MM BREAKAGES. A REVIEW OF THE DEVICE HISTORY FOR THE REPORTED LOT DID NOT INDICATE ANY ABNORMALITIES. A REVIEW OF THE LABELING DID NOT INDICATE ANY ABNORMALITIES. NO INDICATIONS OF MATERIAL OR MANUFACTURING RELATED PROBLEMS WERE FOUND DURING THE INVESTIGATION. IF THE DEVICE IS RETURNED OR IF ANY ADDITIONAL INFORMATION IS PROVIDED, THE INVESTIGATION WILL BE REASSESSED. DEVICE WAS LOST.

Additional Manufacturer Narrative · 1

ONCE THE INVESTIGATION HAS BEEN COMPLETED ANY ADDITIONAL INFORMATION WILL BE REPORTED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

APEX PIN BROKE ON INSERTION INTO TIBIA DURING OPERATION ON RIGHT LEG OF PATIENT.

Description of Event or Problem · 1

APEX PIN BROKE ON INSERTION INTO TIBIA DURING OPERATION ON RIGHT LEG OF PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
713330 SELF-DRILLING HALF PIN APEX Ø 3MM, 110 X 25MM PIN, FIXATION, THREADED JDW STRYKER GMBH F38950

Patients

Seq Age Sex Outcome Treatment
1 58 YR Other