FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 5182072 · Received October 27, 2015

Report

Report Number
2024168-2015-06408
Event Type
Injury
Date Received
October 27, 2015
Date of Event
October 5, 2015
Report Date
November 9, 2015
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). IT WAS REPORTED THAT THE PROGLIDE DEVICE WAS USED IN A CALCIFIED VESSEL. PER THE INSTRUCTIONS FOR USE, THE SAFETY AND EFFECTIVENESS OF THE PROGLIDE DEVICE HAVE NOT BEEN ESTABLISHED FOR PATIENTS WITH FEMORAL ARTERY CALCIUM WHICH IS FLUOROSCOPICALLY VISIBLE AT THE ACCESS SITE. THE DEVICE HAS BEEN RECEIVED. INVESTIGATION IS NOT COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED FOR ANALYSIS. A CUFF MISS WAS ABLE TO BE CONFIRMED. BASED ON A VISUAL AND FUNCTIONAL INSPECTION OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE. THE INVESTIGATION DETERMINED THE DIFFICULTIES AND ADDITIONAL TREATMENT APPEAR TO BE RELATED TO CIRCUMSTANCES OF THE PROCEDURE. A REVIEW OF THE LOT HISTORY RECORD IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. ADDITIONALLY, A REVIEW OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A LOT SPECIFIC QUALITY ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE.

Additional Manufacturer Narrative · 1

(B)(4). DATE OF THIS REPORT - CHANGED FROM NA TO 11/09/2015.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF A MILDLY TORTUOUS, MILDLY CALCIFIED, RIGHT COMMON FEMORAL ARTERY WAS ATTEMPTED USING A PROGLIDE DEVICE WITH A 7F SHEATH AFTER A INTERVENTIONAL PERIPHERAL ARTERY OCCLUSIVE DISEASE (PAOD) PROCEDURE. REPORTEDLY, AN ANTERIOR CUFF MISS OCCURRED. A SECOND PROGLIDE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. THE PHYSICIAN IS REPORTEDLY TRAINED IN THE USE OF THE PROGLIDE DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
712605 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 50429K1

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention