SIG FEM ADPT TORQUE WRENCH
Report
- Report Number
- 1818910-2015-33759
- Event Type
- Malfunction
- Date Received
- October 27, 2015
- Date of Event
- October 23, 2015
- Report Date
- October 23, 2015
- Manufacturer
- DEPUY ORTHOPAEDICS, INC.
- Product Code
- LXH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
(B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
EXAMINATION OF THE SUBMITTED SIG FEM ADPT TORQUE WRENCH CONFIRMS THE BLACK CAP PROTECTOR HAS DISASSEMBLED FROM THE WRENCH. CAPA 000509 WAS INITIATED TO FURTHER INVESTIGATE AND DETERMINE ROOT CAUSE AND CORRECTIVE ACTIONS. CO 103025887 WAS RELEASED ON JANUARY 2015 TO CHANGE THE DESIGN OF PRODUCT CODE 961673 AS PART OF CAPA 000509. THE CURRENT COMPLAINT SAMPLE WAS MANUFACTURED PRIOR TO THE IMPLEMENTATION OF CO 103025887. NO FURTHER CORRECTIVE ACTION REQUIRED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
THE PLASTIC/RUBBER TIP IS MISSING FROM THE END.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 711096 | SIG FEM ADPT TORQUE WRENCH | KNEE INSTRUMENT/TRIAL | LXH | DEPUY ORTHOPAEDICS, INC. | SO2012916 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |