FDA Adverse Event Malfunction Summary report: N

REPLY

MDR report key: 5179833 · Received October 27, 2015

Report

Report Number
1000165971-2015-00636
Event Type
Malfunction
Date Received
October 27, 2015
Date of Event
October 1, 2015
Report Date
October 1, 2015
Manufacturer
SORIN GROUP ITALIA S.R.L. - CRM FACILITY
Product Code
LWP
PMA / PMN Number
PP950029
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

PLEASE REFER TO THE ATTACHED ANALYSIS REPORT.

Description of Event or Problem · 1

UPON INTERROGATION OF THE PACEMAKER ON (B)(6) 2015, THE LANGUAGE WAS (B)(6) INSTEAD OF ENGLISH. MOREOVER, THE PACEMAKER WAS FOUND SWITCHED IN STAND-BY MODE. AS-SHIPPED SETTINGS WERE PROGRAMMED AFTER RE-INITIALIZATION OF THE PACEMAKER, EXCEPT FOR THE POLARITIES THAT WERE SET TO UNIPOLAR. REPORTEDLY, VENTRICULAR LEAD IMPEDANCE VALUE WAS ALSO FLUCTUATING. PRELIMINARY ANALYSIS SHOWED THAT THE PACEMAKER SWITCHED IN STANDBY MODE ON (B)(6) 2015, MOST PROBABLY DUE TO A SINGLE EVENT UPSET (SEU). THE PACEMAKER WAS PROPERLY REINITIALIZED ON (B)(6) 2015 AND ITS NORMAL FUNCTIONING WAS RESTORED. IT ALSO APPEARS THAT THE LANGUAGE OF THE PROGRAMMER SWITCHED TO GERMAN AFTER USING THE PACEART EXPORT FUNCTION.

Description of Event or Problem · 1

UPON INTERROGATION OF THE PACEMAKER ON (B)(6) 2015, THE LANGUAGE WAS (B)(6) INSTEAD OF ENGLISH. MOREOVER, THE PACEMAKER WAS FOUND SWITCHED IN STANDBY MODE. AS-SHIPPED SETTINGS WERE PROGRAMMED AFTER RE-INITIALIZATION OF THE PACEMAKER, EXCEPT FOR THE POLARITIES THAT WERE SET TO UNIPOLAR. REPORTEDLY, VENTRICULAR LEAD IMPEDANCE VALUE WAS ALSO FLUCTUATING. PRELIMINARY ANALYSIS SHOWED THAT THE PACEMAKER SWITCHED IN STANDBY MODE ON (B)(6) 2015, MOST PROBABLY DUE TO A SINGLE EVENT UPSET (SEU). THE PACEMAKER WAS PROPERLY REINITIALIZED ON (B)(6) 2015 AND ITS NORMAL FUNCTIONING WAS RESTORED. IT ALSO APPEARS THAT THE LANGUAGE OF THE PROGRAMMER SWITCHED TO GERMAN AFTER USING THE PACEART EXPORT FUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
711430 REPLY PULSE GENERATOR, PERMANENT, IMPLANTABLE LWP SORIN GROUP ITALIA S.R.L. - CRM FACILITY REPLY DR 2404

Patients

Seq Age Sex Outcome Treatment
1