FDA Adverse Event Malfunction Summary report: N

RNS-9703-024

MDR report key: 5176296 · Received October 25, 2015

Report

Report Number
2032233-2015-00327
Event Type
Malfunction
Date Received
October 25, 2015
Date of Event
September 25, 2015
Report Date
September 25, 2015
Manufacturer
NIHON KOHDEN CORPORATION
Product Code
MHX
PMA / PMN Number
K023475
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

NIHON KOHDEN CONTINUES TO INVESTIGATE THE REPORTED EVENT. NIHON KOHDEN WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR PART 803.56 WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

THE CUSTOMER REPORTED THAT THE RNS (REMOTE NETWORK SYSTEM OR REMOTE MONITORING SYSTEM) IS CRASHING WITH ERROR CODE "THE MEMORY COULD NOT BE READ" BY CNS6201.EXE. THE BIOMEDICAL ENGINEER ATTEMPTED TO REBOOT THE MACHINE WHICH STARTS THE PROGRAM AGAIN BUT THEN IT CRASHED AGAIN. THE DEVICE WAS EVALUATED AND THE REPORTED PROBLEM WAS DUPLICATED. THE CUSTOMER IDENTIFIED THE CAUSE OF THE MALFUNCTION. A REPLACEMENT MODULE OR COMPONENT WAS SENT TO THE CUSTOMER. BASED ON THE INFORMATION PROVIDED AT THE TIME OF THE EVENT ASSESSMENT, THERE IS NO INDICATION THAT WAS A PATIENT INJURY OR A REPORTABLE EVENT AS A RESULT OF THIS ISSUE. NIHON KOHDEN WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR PART 803.56 IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE RNS (REMOTE NETWORK SYSTEM OR REMOTE MONITORING SYSTEM) IS CRASHING WITH ERROR CODE "THE MEMORY COULD NOT BE READ" BY CNS6201.EXE. THE BIOMEDICAL ENGINEER ATTEMPTED TO REBOOT THE MACHINE WHICH STARTS THE PROGRAM AGAIN BUT THEN IT CRASHED AGAIN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
707094 RNS-9703-024 REMOTE NETWORK STATION MHX NIHON KOHDEN CORPORATION RNS-9703-024

Patients

Seq Age Sex Outcome Treatment
1