BONE PIN- 3.2MM X 140MM- STERILE 2 PACK
Report
- Report Number
- 3005985723-2015-00213
- Event Type
- Malfunction
- Date Received
- October 24, 2015
- Date of Event
- October 21, 2015
- Report Date
- October 21, 2015
- Manufacturer
- MAKO SURGICAL CORP.
- Product Code
- OLO
- PMA / PMN Number
- K112507
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
FOLLOW-UP #1. REPORTED EVENT: THE 3.2 X 140 BONE PIN BROKE WHILE SURGEON WAS REMOVING IT FROM PATIENT'S TIBIA. IT BROKE OFF BELOW THE SURFACE OF THE SKIN AND WAS LEFT IN PATIENT. MINIMAL DELAY TO CASE, NO ADVERSE AFFECTS TO PATIENT, AND CASE WAS SUCCESSFULLY FINISHED. DEVICE EVALUATION AND RESULTS: VISUAL INSPECTION WAS NOT PERFORMED AS THE DEVICE WAS NOT RETURNED FOR EVALUATION. DIMENSION INSPECTION COMPLETED AT THE TIME OF MANUFACTURING SHOWS THE LOT WAS WITHIN SPECIFICATION. DEVICE HISTORY REVIEW: REVIEW OF THE DEVICE HISTORY RECORDS FOR THE ASSOCIATED LOT INDICATED (B)(4) DEVICES ((B)(4) NON-STERILE BONE PIN, SINGLE) WERE MANUFACTURED AND SHIPPED TO MILLSTONE ON 07/23/2015 AND ACCEPTED INTO FINAL STOCK ON 07/23/2015 PER ERP. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS IN TRACKWISE AND CATSWEB RELATED TO P/N 143140, LOT NUMBER W41379-2 SHOWS NO ADDITIONAL COMPLAINTS RELATED TO THE FAILURE IN THIS INVESTIGATION. TRACKING OF COMPLAINTS RELATED TO THE 1143140 PART NUMBER WILL BE TRACKED THROUGH QUARTERLY TREND REQUEST (B)(4). CONCLUSIONS: AS PART OF THE INVESTIGATION INTO THIS COMPLAINT, THE STRYKER REP THAT WAS PRESENT FOR THE CASE WAS CONTACTED. IT WAS DETERMINED THAT THE SURGEON DID NOT USE THE DRILL GUIDE DURING PLACEMENT OF BOTH BONE PINS. DUE TO THE NON-USE OF A DRILL GUIDE AS REQUIRED PER MAKOPLASTY PARTIAL KNEE APPLICATION USER GUIDE, (B)(4) REVISION 01 FOR PLACEMENT OF BONE PINS, THIS EVENT CONSTITUTES AN OFF-LABEL USE OF THE DEVICE. CORRECTIVE ACTION/PREVENTIVE ACTION: NO ACTION IS REQUIRED AT THIS TIME AS THERE IS NO INDICATION TO SUGGEST A PRODUCT NON-CONFORMITY OR UNANTICIPATED HAZARD.
AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.
THE SURGEON PERFORMED A PARTIAL KNEE ARTHROPLASTY PROCEDURE USING THE ROBOTIC ARM INTERACTIVE ORTHOPEDIC SYSTEM (RIO). DURING THE CASE, THE TIPS OF THE BONE PINS BROKE IN THE PATIENT'S BONE. THE SURGEON DECIDED TO LEAVE THE TIP OF THE BONE PINS IN THE PATIENT'S BONE. THE OUTCOME OF THE CASE WAS SUCCESSFUL.
THE SURGEON PERFORMED A PARTIAL KNEE ARTHROPLASTY PROCEDURE USING THE ROBOTIC ARM INTERACTIVE ORTHOPEDIC SYSTEM (RIO). DURING THE CASE, THE TIPS OF THE BONE PINS BROKE IN THE PATIENT'S BONE. THE SURGEON DECIDED TO LEAVE THE TIP OF THE BONE PINS IN THE PATIENT'S BONE. THE OUTCOME OF THE CASE WAS SUCCESSFUL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 704541 | BONE PIN- 3.2MM X 140MM- STERILE 2 PACK | STEREOTACTIC DEVICE, ACCESSORY | OLO | MAKO SURGICAL CORP. | W41379-2 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Other |